Skip to main content

Effects of blood collection tubes, including pediatric devices, on 16 common immunoassays.

Wang S, Ho V, Roquemore-Goins A, Smith FA

Publication: Clin Chem (2006) • Vol 52 • Page 892-3PMID: 16638955View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection tube type on 16 clinical chemistry analytes in blood.

Conclusion

11 of 16 clinical chemistry analytes were not significantly different between specimens collected in plain glass Vacutainers (control) and Vacutainers or Microtainers containing cell separator, gel separator and/or lithium heparin. Free triiodothyronine (FT3) and free thyroxine (FT4) were increased by collection in tubes with gel separator. Triiodothyronine (T3) and thyroxine (T4) were decreased, and parathyroid hormone (PTH) was increased by collection in tubes containing heparin. There were no differences detected in any of the 16 analytes between plain glass Vacutainers and modified plain Microtainers.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection tube on 16 clinical chemistry analytes in blood. In all cases, Vacutainer tubes were filled before Microtainer tubes.

Summary of Findings

11 of 16 clinical chemistry analytes were not significantly different between specimens collected in plain glass Vacutainers (control) and Vacutainers or Microtainers containing cell separator, gel separator and/or lithium heparin. FT3 and FT4 were increased by collection in tubes with gel separator. T3 and T4 were decreased, and PTH was increased by collection in tubes containing heparin. There were no differences detected in any of the 16 analytes between plain glass Vacutainers and modified plain Microtainers.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    GlycoproteinClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Glass Vacutainer (plain)
    Microtainer (modified, plain)
    Microtainer with lithium heparin and gel separator
    Microtainer with serum cell separator (modified)
    Plastic Vacutainer with lithium heparin
    Plastic Vacutainer with lithium heparin and gel separator
    Plastic Vacutainer with serum cell separator (modified)