Surface-enhanced laser desorption ionization/time-of-flight mass spectrometry reveals significant artifacts in serum obtained from clot activator-containing collection devices.

Pilny R, Bouchal P, Borilova S, Ceskova P, Zaloudik J, Vyzula R, Vojtesek B, Valik D

Publication: Clin Chem (2006) • Vol 52 • Page 2115-6PMID: 17068172View on PubMed

Purpose

The purpose of this paper was to determine the effects of collecting blood in clot activator-containing tubes and frozen storage of serum on proteomic analysis.

Conclusion

Mass spectrometry (MS) spectra of specimens collected in tubes containing clot activator had two peaks with >40-fold higher peak intensities than those seen for specimens collected in tubes without clot activator. Further, these two peaks were next to two other peaks which were decreased in some spectra from specimens collected in tubes with clot activator. No changes in spectra occurred during two months of additional frozen storage.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collecting blood in clot activator-containing tubes and frozen storage of serum on proteomic analysis. All serum was stored in aliquots at -25 degrees C prior to analysis. A second set of 10 aliquots was analyzed 2 months after the first set.

Summary of Findings

MS spectra of specimens collected in tubes containing clot activator had two peaks with >40-fold higher peak intensities than those seen for specimens collected in tubes without clot activator. Further, these two peaks were next to two other peaks which were decreased in some spectra from specimens collected in tubes with clot activator. No changes in spectra occurred during two months of additional frozen storage.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    PeptideSELDI-TOF MS
    ProteinSELDI-TOF MS
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Microvette Sarstedt Serum Gel Clotting activator tubes
    Microvette Sarstedt tubes
    StorageStorage duration
    Unspecified
    Unspecified + 2 months