Elevated results in a vitamin B12 assay when using serum separator blood collection tubes.
Lowrey I, Smith G
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of using SST versus PT, specimen mixing, re-centrifugation, and tube cap type on vitamin B12 measurement in serum.
Summary of Findings
Average vitamin B12 was significantly higher in serum collected in SST (256 pmol/L) than PT (240 pmol/L). 150 min of room temperature mixing in the original SST after centrifugation caused a significant increase in serum vitamin B12 (394 pmol/L), and this increase was attenuated, although not abolished, by recentrifugation of the specimens. The authors state that using 4 mL capacity SST as opposed to the standard 9.5 mL SST in 3 cases resulted in similar increases in vitamin B12 after mixing, but unlike in the larger SST, these increases were completely abolished by recentrifugation. Neither the type of cap on the tubes during mixing (unused SST cap or new PT cap) nor an additional 150 min of mixing (300 min versus 150 min) had any effects on serum vitamin B12 concentrations.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Small moleculeChemiluminescencePre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution 4 mL SST7 mL PT9.5 mL SSTBiospecimen Aliquots and Components Biospecimen mixing 150 min at room temperature300 min at room temperatureNoneBiospecimen Aliquots and Components Centrifugation Different number of centrifugation steps comparedStorage Type of storage container Cap from unused PTCap from unused SST