Skip to main content

A comparison of glass and plastic blood collection tubes for routine and specialized coagulation assays: a comprehensive study.

Kratz A, Stanganelli N, Van Cott EM

Publication: Arch Pathol Lab Med (2006) • Vol 130 • Page 39-44PMID: 16390236View on PubMed

Purpose

The purpose of this paper was to determine the effects of using plastic versus glass blood collection tubes on 22 different coagulation assays in plasma.

Conclusion

14 of the 22 coagulation assays showed no statistical differences between specimens collected in glass BD Vacutainer tubes and plastic BD Vacutainer Plus tubes. The tests that did show a statistically significant difference between tube types had average differences of 0.4%-5.5%, and the authors state they are unlikely to be clinically significant.

Studies (1)

Purpose

The purpose of this study was to determine the effects of using plastic versus glass blood collection tubes on 22 different coagulation assays in plasma. Some specimens were analyzed immediately after processing while others were stored at -70 degrees C until analysis.

Summary of Findings

14 of the 22 coagulation assays showed no statistical differences between specimens collected in glass BD Vacutainer tubes and plastic BD Vacutainer Plus tubes. The tests that did show a statistically significant difference between tube types included fibrinogen, protein C activity, Factor X, Factor XI, and Factor XII which were increased in plastic tubes compared to glass tubes, and protein S activity, lupus anticoagulant, and Factor VII which were decreased in plastic tubes compared to glass tubes. These assays had average differences of 0.4%-5.5% between tube types, and the authors state they are unlikely to be clinically significant.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)Frozen
  • Diagnoses

    Normal
  • Platforms

    ProteinHematology/ auto analyzer
    MorphologyHematology/ auto analyzer
    GlycoproteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Glass Vacutainer tubes
    Plastic Vacutainer Plus tubes
    Biospecimen PreservationType of fixation/preservation
    Frozen
    None (fresh)