Comparison of an electrochemical biosensor with optical devices for hemoglobin measurement in human whole blood samples.
Hsieh MS, Wu TG, Su CS, Cheng WJ, Ozbek N, Tsai KY, Lin CY
Purpose
The purpose of this paper was to determine the effects of delayed analysis, biospecimen components and analyzer type on the measurement of hemoglobin in potassium EDTA blood.
Conclusion
Storage for 7 days at 4 degrees or room temperature, hemolysis, and bilirubin did not interfere with hemoglobin measurement in blood specimens.
Studies (1)
Purpose
The purpose of this study was to determine the effects of delayed analysis, biospecimen components, hemolysis, and analyzer type on the measurement of hemoglobin in blood.
Summary of Findings
Storage of blood specimens for 7 days at either 4 degrees C or room temperature had no effects on the measurement of hemoglobin. Hemolysis caused by the addition of hemolysis reagent, or freeze-thaw cycling and vortexing caused a less than 6% bias in hemoglobin measurement. Further, added bilirubin did not interfere with hemoglobin measurement. The results obtained using the BeneCheck analyzer were highly correlated with those obtained with Sysmex, Beckman Coulters, Cell-Dyn, and HemoCue analyzers, and the variance between results was generally less than 5%.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components 0-150 mg/L bilirubin110-150 g/L hemoglobinBiospecimen Aliquots and Components Hemolysis Chemically-inducedFreeze/thaw-inducedNot inducedVortex-inducedStorage Storage duration 0 days1 day3 days7 daysStorage Storage temperature 4 degrees CRoom temperatureHematology/ auto analyzer Specific Technology platform Beckman CoultersBeneCheckCell-DynHemoCueSysmex