Influence of serum separator tubes on total and free phenytoin concentrations and dosages.
Cai WM, Leader WG, Porter WH, Chandler MH
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of using SST instead of plain tubes for the collection and storage of serum on the measurement of total and free phenytoin and albumin concentrations. Specimens were kept at room temperature for the first 24 h of storage, after which point they were refrigerated at 4 degrees C.
Summary of Findings
Compared to total phenytoin concentrations measured in serum collected in plain tubes, concentrations measured in serum from SST were significantly lower 2 h after collection and continued to decrease during 24 h at room temperature and an additional 24 h at 4 degrees C (48 h total storage) when centrifuged serum was kept in the collection tubes. In plain tubes, total phenytoin concentrations remained stable throughout the 48 h of storage. Free phenytoin concentrations were significantly lower in SST than in plain tubes after 48 h, but not 2 or 24 h of storage. The free fraction of unbound phenytoin was significantly higher in SST than in plain tubes after 24 h of storage. Serum albumin levels were not significantly different between the two tube types after 24 h of storage.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Other PreservativeDiagnoses
Not specifiedPlatforms
Small moleculeClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Plain tubeSSTStorage Storage duration 2 h24 h48 h