Lupus anticoagulant testing: analyzing fresh samples after a single centrifugation and after a 6-8 h delay.

Froom P, Barak M

Publication: Clin Chem Lab Med (2011) • Vol 50 • Page 367-370PMID: 22040237View on PubMed

Purpose

The purpose of this paper was to determine the effects of a second centrifugation step and storage of plasma on lupus anticoagulant testing.

Conclusion

A second centrifugation step did not change the dilute Russell viper venom test (dRVVT) or silica clotting time (SCT) outcomes when plasma was undergoing lupus anticoagulant testing. Further, the dRVVT and SCT results were not significantly changed when testing was performed 6-8 h after specimen collection rather than less than 4 h after.

Studies (1)

Purpose

The purpose of this study was to determine the effects of a second centrifugation step and storage of plasma on lupus anticoagulant testing. The temperature used during storage was not reported.

Summary of Findings

A second centrifugation step did not change the dRVVT or SCT results for plasma. Further the dRVVT and SCT results were not significantly changed when lupus anticoagulant testing was performed 6-8 h after specimen collection rather than less than 4 h after..

  • Biospecimens

    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Lupus
    Normal
  • Platforms

    MorphologyHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCentrifugation
    Different number of centrifugation steps compared
    StorageStorage duration
    6-8 h
    <4 h
    Hematology/ auto analyzer SpecificTechnology platform
    SCT
    dRVVT