Skip to main content

Comparison of plastic vs. glass evacuated serum-separator (SST) blood-drawing tubes for common clinical chemistry determinations.

Hill BM, Laessig RH, Koch DD, Hassemer DJ

Publication: Clin Chem (1992) • Vol 38 • Page 1474-8PMID: 1643717View on PubMed

Purpose

The purpose of this paper was to determine the effects of using glass versus plastic serum separator tubes (SST) for collection and storage on the measurement of common clinical chemistry analytes in serum.

Conclusion

In general, there were no differences in the measurement of common clinical chemistry analytes or the detection of drugs in serum collected and stored in glass versus plastic SST.

Studies (1)

Purpose

The purpose of this study was to determine the effects of using glass versus plastic SST for collection and for 48 hour of refrigerated storage on the measurement of 32 common clinical chemistry analytes and the detection of 6 drugs in serum using different auto analyzers.

Summary of Findings

Of 226 different analyte-instrument combinations evaluated, the authors report that only lactate dehydrogenase (LD) measured on the Technicon SMAC was significantly different between plastic and glass SST (p=0.01), while two other auto analyzers (Ektachem 700 and Hitachi 737) did not detect a difference in LD between the collection tube types. When serum was allowed to remain on the gel barrier for 48 hours of refrigerated storage, only carbon dioxide (CO2) was significantly different between glass and plastic SST, the later having a slightly lower average level. However, there were no significant differences in CO2 between glass and plastic SST tubes when serum was analyzed immediately or between CO2 measured immediately versus after 48 h storage in plastic SST.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)Other Preservative
  • Diagnoses

    Not specified
    Normal
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Glass Vacutainer SST
    Plastic Vacutainer Plus SST
    Biospecimen PreservationType of fixation/preservation
    None (fresh)
    Refrigeration
    StorageStorage duration
    0 h
    48 h
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Dimension 360
    Ektachem 700
    Hitachi 737
    SMAC
    TDx
    aca