Clinically relevant differences in prothrombin time and INR values related to blood sample collection in plastic vs glass tubes.
Fiebig EW, Etzell JE, Ng VL
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube type on PT and INR in platelet-poor plasma. Plasma used for comparison of auto analyzers was stored at -70 degrees C for less than 20 days, on dry ice for 1 h, and then at -20 degrees C for less than 48 h or at -70 degrees C for the entire time prior to analysis.
Summary of Findings
Specimens collected in plastic Vacuette or Vacutainer Plus tubes, as opposed to glass Monoject tubes, had significantly shorter PTs, and specimens with higher PTs showed the largest decreases (16-18%) compared to specimens in glass tubes. PTs measured in specimens collected in either type of plastic tube were similar. INR values were lower in specimens collected in either type of plastic tube versus glass tubes. The duration of warfarin therapy did not influence the differences in INRs observed between plastic and glass tubes, nor did the auto analyzer/reagent combination used for analysis.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)FrozenDiagnoses
Not specifiedPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Other drugs Unspecified/noneWarfarin therapy <6 weeksWarfarin therapy >6 weeksBiospecimen Acquisition Type of collection container/solution 2.7 mL plastic Vacutainer Plus3.5 mL plastic Vacuette4.5 mL glass MonojectBiospecimen Preservation Type of fixation/preservation FrozenNone (fresh)Hematology/ auto analyzer Specific Reaction solution Thrombomax HSThromboplastin-DSHematology/ auto analyzer Specific Technology platform AMAX 190MLA Electra 1400C analyzer