Effects of storage and type of blood collection tubes on hepatitis C virus level in whole blood samples.
Kessler HH, Stelzl E, Raggam RB, Haas J, Kirchmeir F, Hegenbarth K, Daghofer E, Santner BI, Marth E, Stauber RE
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of room temperature storage and collection tube type on HCV RNA levels in whole blood. HCV RNA levels were also compared between serum and whole blood. Serum and whole blood specimens were frozen at -70 degrees C after experimental room temperature storage.
Summary of Findings
Whole blood collected in EDTA tubes had smaller coefficients of variation and yielded higher concentrations of HCV RNA than whole blood collected in nucleic acid stabilization tubes (NASTs). After 48 h of room temperature storage, HCV RNA concentrations remained unchanged in all whole blood collection tubes, however, after 96 h, only HCV RNA measured in blood collected in Vacuette NASTs remained stable while significant increases were observed in blood collected in Vacuette EDTA (p<0.001) and Vacutainer EDTA tubes (p<0.007). Mean HCV RNA concentrations were comparable between serum and whole blood collected in NASTs, but concentrations were higher in whole blood collected in either Vacuette or Vacutainer EDTA tubes than in serum.
Biospecimens
FluidSerumFluidBloodPreservation Types
FrozenDiagnoses
HepatitisPlatforms
RNAReal-time qRT-PCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution 3.0 mL Vacuette EDTA3.0 mL Vacutainer EDTA3.5 mL Vacuette NAST9.0 mL VacutainerBiospecimen Aliquots and Components Blood and blood products SerumWhole bloodStorage Time at room temperature 0 h48 h96 h