Preanalytical variables and off-site blood collection: influences on the results of the prothrombin time/international normalized ratio test and implications for monitoring of oral anticoagulant therapy.
van Geest-Daalderop JH, Mulder AB, Boonman-de Winter LJ, Hoekstra MM, van den Besselaar AM
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of delayed centrifugation, centrifugation conditions (room temperature or 20°C), centrifugation duration (5 min or 10 min) and storage of plasma after centrifugation on PT in plasma. The study was conducted at 4 different centers, each using different collection tubes types and analysis methods.
Summary of Findings
When tubes of whole blood were stored prior to centrifugation at 4 degrees C, room temperature, or 37 degrees C in center 2 which used Venoject tubes and analyzed with Recombiplastin the PT decreased progressively with storage, except for in tubes stored at 37 degrees C where the PT increased between 3-6 h and 6-24 h. In contrast, when centrifugation was delayed at Center 1 which used Monovette or plastic Vacutainer tubes and analyzed PT using HepatoQuick, no changes were observed in PT of specimens stored at room temperature, but PT increased in specimens stored for 24 h at 4 or 37 degrees C and decreased in specimens stored for 3 h at 37 degrees C. Interestingly, when specimens were stored at room temperature with agitation prior to centrifugation, all specimens, except those from center 2 stored for 3 h, showed significant increases in PT. In specimens from center 3, which used glass Vacutainers and analyzed PT with Symplastin, storage of blood with agitation prior to centrifugation resulted in progressive decreases in PT at 2-4 degrees C, a decrease in PT after 3 h at 37 degrees C, and an increase in PT after 24 h at 37 degrees C. After centrifugation, storage of plasma for up to 24 h prior to analysis using the conditions of center 2 had no effect on PT, but using center 1's conditions, PT was decreased when centrifuged specimens were stored for 3 or 6, but not 24 h prior to analysis. There were no significant differences between the PT of specimens centrifuged at a controlled temperature of 20 degrees C and those centrifuged without temperature control, and centrifugation time also had no effects on PT.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Blood and blood products PlasmaWhole bloodBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedMultiple durations comparedMultiple temperatures comparedStorage Storage conditions Mechanical agitationNo agitationStorage Storage duration 0.5-1 h24 h3 h6 hStorage Storage temperature 37 degrees C4-6 degrees CRoom temperatureStorage Type of storage container Glass vacutainerMonovette or plastic vacutainerVenojectHematology/ auto analyzer Specific Technology platform HepatoQuickRecombiplastinSymplastin