The use of Vacutainer tubes for collection and storage of blood for coagulation testing.
Thomson JM, Easton AC, Faragher EB
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of storage container, storage temperature and patient condition on coagulation parameters. Specimens were first collected through a syringe and distributed into polystyrene tubes containing trisodium citrate, and subsequent specimens were collected directly into vacutainers.
Summary of Findings
For healthy individuals, the PT decreased after 2 to 4 h of storage at 4 degrees C or room temperature in specimens held in vacutainers, but PT was not affected by storage of blood in polystyrene tubes, regardless of storage temperature. Similarly, in patients on coumarin therapy there was a decrease in PT in specimens stored in vacutainers for 4 h at 4 degrees C, but not at room temperature, compared to specimens analyzed after 0 or 2 h or specimens stored in polystyrene tubes. In specimens obtained from women in the third trimester of pregnancy, PT was decreased when blood was stored for 4 h in polystyrene tubes or vacutainers at 4 degrees C but not when specimens were stored at room temperature. When blood from pregnant women was stored at 4 degrees C in vacutainers, Factor VII increased in specimens that were stored for 2 h or more. In specimens from healthy individuals and patients on coumarin therapy, Factor VII increased after 4 h of storage at 4 degrees C. In specimens from healthy individuals and pregnant women, APTT was lengthened in specimens stored at 4 degrees C in vacutainers or polystyrene tubes for 4 h, but APTT was not affected by storage at room temperature or storage of specimens from patients on coumarin therapy.
Biospecimens
FluidBloodPreservation Types
FrozenDiagnoses
NormalOther diagnosesPregnantPlatforms
MorphologyHematology/ auto analyzerProteinHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition HealthyPatients on coumarinThird-trimester of pregnancyStorage Storage duration 0 h2 h4 hStorage Storage temperature 4 degrees CRoom temperatureStorage Type of storage container Polystyrene tube with trisodium citrateVacutainer