Effect of different sample types and stability after blood collection of N-terminal pro-B-type natriuretic peptide as measured with Roche Elecsys system.
van der Merwe DE, Henley R, Lane G, Field R, Frenneaux M, Dunstan F, McDowell I
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube type and storage in SST on the measurement of NT-proBNP. Serum and plasma specimens were frozen at -20 degrees C until analysis.
Summary of Findings
NT-proBNP was 4% lower in lithium-heparin plasma than in serum (p<0.001), but there were no differences in NT-proBNP between serum collected in SST and glass tubes. NT-proBNP decreased in serum with increasing storage duration when whole blood was stored unseparated at room temperature in an SST, but when separated serum was stored at room temperature in an SST, the NT-proBNP levels remained stable for at least 72 h.
Biospecimens
FluidBloodFluidPlasmaFluidSerumPreservation Types
None (Fresh)Diagnoses
Cardiovascular DiseaseNot specifiedPlatforms
PeptideClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Lithium heparin tubePlain glass tubeSST tubeBiospecimen Aliquots and Components Blood and blood products PlasmaSerumWhole bloodBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Time at room temperature 0 h24 h48 h72 h