Comparison between siliconized evacuated glass and plastic blood collection tubes for prothrombin time and activated partial thromboplastin time assay in normal patients, patients on oral anticoagulant therapy and patients with unfractioned heparin therapy.
D'Angelo G, Villa C
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube type on PT and aPTT in platelet poor plasma from normal individuals and those on oral anticoagulant therapy.
Summary of Findings
In platelet poor plasma from normal patients and those being treated with unfractionated heparin, aPTT test results were not statistically different between collection tube types. In normal patients and those being treated with oral anticoagulant therapy with vitamin K antagonists, PT was longer when platelet poor plasma was collected in Kima and Greiner plastic tubes than when collected in BD glass tubes or Terumo plastic tubes. There was also a high correlation between PT in Terumo and BD tubes (r2=0.97).
Biospecimens
FluidBloodFluidPlasmaPreservation Types
None (Fresh)Diagnoses
NormalNot specifiedPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Other drugs No oral anticoagulant therapyUnfractionated heparinVitamin K antagonistsBiospecimen Acquisition Type of collection container/solution BD glassGreiner plasticKima plasticTerumo plastic