Impact of analytical bias in metabonomic studies of human blood serum and plasma.
Teahan O, Gamble S, Holmes E, Waxman J, Nicholson JK, Bevan C, Keun HC
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of clotting time and temperature, freeze-thaw cycling, use of plasma instead of serum and delayed analysis on NMR spectra. After experimental storage, plasma and serum specimens were stored at -80 degrees C until analysis.
Summary of Findings
More than 50% of the variation in NMR spectra in serum was attributed to differences between the 4 patients. Clotting time increased the peaks associated with lactate, but this was attenuated when clotting was performed on ice rather than at room temperature. Only small differences in spectra were observed between heparinated plasma and serum specimens, and these were predominantly attributed to lipids. Further, small differences in average peak height were observed after serum from 24 individuals was subjected to an additional freeze-thaw cycle, but they were 2 orders of magnitude less than the average peak height and were predominantly due to changes in high molecular weight species. When 1 sample was stored in the NMR machine prior to analysis, a less than 5% variation in spectra was observed over the course of 3 h, but the authors state the variations in lipids that occurred might cause a systematic bias if analysis is delayed.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
CarbohydrateNMRLipidNMRSmall moleculeNMRSteroidNMRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Blood and blood products PlasmaSerumBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Freeze/thaw cycling 1 cycle2 cyclesStorage Storage duration 120 min180 min30 min60 minStorage Storage temperature On iceRoom temperature