Controlled storage conditions prolong stability of biochemical components in whole blood.
Stahl M, Brandslund I
Purpose
Conclusion
Studies (2)
Purpose
The purpose of this study was to determine the effects of centrifugation at 1200 x g versus 3000 x g on 7 clinical chemistry analytes in plasma. After centrifugation, plasma was refrigerated at 4 degrees C until analysis.
Summary of Findings
Centrifugation speed had no effects on albumin, calcium, glucose, lactate dehydrogenase, potassium, sodium or total protein in plasma.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Centrifugation Multiple speeds compared
Purpose
The purpose of this study was to determine the effects of delayed centrifugation on 27 clinical chemistry analytes in plasma. After centrifugation, plasma was stored at 4 degrees C until analysis.
Summary of Findings
Potassium increased with increasing delays to centrifugation at 17 or 20 degrees C and decreased with increasing delays to centrifugation at 23 or 25 degrees C. However, when specimens were held at 20 degrees C, prior to centrifugation, the change in potassium was clinically irrelevant until at least an 8 h delay. Plasma phosphate decreased when blood was stored at 20 degrees C for 8 h or more before centrifugation. No other clinically relevant effects of delayed centrifugation were observed.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerGlycoproteinClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerPeptideClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Storage duration 0 h10 h12 h2 h4 h6 h8 hStorage Storage temperature 17 degrees C20 degrees C23 degrees C25 degrees C