Plasma creatinine and the effect of delay in separation of samples.
Simpson WG
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of delayed centrifugation of whole blood on creatinine concentrations in lithium heparin-plasma using the kinetic Jaffe method on an autoanalyzer.
Summary of Findings
Creatinine concentrations increased in lithium heparin-plasma when centrifugation was delayed by as little as 8 h (p=0.011) and peaked at 48 h, but the increase was clinically relevant only when centrifugation was delayed by more than 24 h. The rate of creatinine increase in all 5 specimens was fastest between 24 and 48 h.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Small moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Time at room temperature 2 h24 h48 h72 h8 h