Leakage of oxygen from blood and water samples stored in plastic and glass syringes.

Scott PV, Horton JN, Mapleson WW

Publication: Br Med J (1971) • Vol 3 • Page 512-6PMID: 5565518View on PubMed

Purpose

The purpose of this paper was to determine the effects of syringe material and initial partial pressure oxygen (PO2) on PO2 of tonometered blood.

Conclusion

When blood tonometered to 680 mm mercury (Hg) oxygen (O2) was stored in plastic syringes for 11.5 min, the PO2 was 10% lower than that measured after blood was stored in glass syringes. After storage of specimens tonometered with 102 or 68 mm Hg O2, PO2 was similar between specimens stored in plastic and glass syringes.

Studies (1)

Purpose

The purpose of this study was to determine the effects of initial PO2 and storing blood for 11.5 min at room temperature in plastic rather than glass syringes on PO2 of tonometered blood. Heparinized whole blood was tonometered with O2 or air at 37 degrees C.

Summary of Findings

When blood tonometered to 680 mm Hg O2 was stored in plastic syringes for 11.5 min, the PO2 was 10% lower than that measured after blood was stored in glass syringes. After storage of specimens tonometered with 102 or 68 mm Hg O2 for 11.5 min, PO2 was similar between specimens stored in plastic and glass syringes. Predicted effects of blood temperature, pH, and hemoglobin on PO2 of blood and the effects of several syringe types and storage temperatures on PO2 of water were also presented.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    GasRadiometer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen components
    102 mm Hg O2
    68 mm Hg O2
    680 mm Hg O2
    StorageType of storage container
    Brand D plastic syringe
    Glass syringe