Stability of creatinine with delayed separation of whole blood and implications for eGFR.

Shepherd J, Warner MH, Kilpatrick ES

Publication: Ann Clin Biochem (2007) • Vol 44 • Page 384-7PMID: 17594786View on PubMed

Purpose

The purpose of this paper was to determine the effects of delayed centrifugation and method of analysis on creatinine measurement.

Conclusion

When centrifugation was delayed by more than 14 h, there was an increase in creatinine levels as measured using the three kinetic Jaffe assays, but creatinine levels were stable after centrifugation delays of up to 31 h when assayed using the two enzymatic assays. The increase in creatinine levels after an 11-28 h delay to centrifugation resulted in decreased estimated glomerular filtration rate (eGFR) which led to a misclassification of chronic kidney disease in 4 of 21 patients.

Studies (1)

Purpose

The purpose of this study was to determine the effects of delayed centrifugation and method of analysis on creatinine measurements in healthy individuals. Blood from 5 individuals was subjected to different delays in centrifugation and analyzed on the 5 different analyzers. Subsequently blood specimens from 24 patients were analyzed within 1 h and after a delay of up to 28 h using the Beckman DXC 800 CRE 3 kinetic Jaffe assay.

Summary of Findings

When centrifugation was delayed by more than 14 h, there was an increase in creatinine levels as measured using the three kinetic Jaffe assays, but creatinine levels were stable after centrifugations delays of up to 31 h when assayed using the two enzymatic assays. The largest increase in measured creatinine concentration after delayed centrifugation occurred with the Beckman DXC 800 CRE 3 kinetic Jaffe assay. The increase in creatinine levels after an 11-28 h delay to centrifugation resulted in decreased eGFR which led to a misclassification of chronic kidney disease in 4 of 21 patients.

  • Biospecimens

    FluidBlood
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
    Other diagnoses
    Normal
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Chronic kidney disease
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageTime at room temperature
    11-28 h
    14 h
    15 min
    24 h
    31 h
    4 h
    8 h
    <1 h
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Bayer Advia kinetic Jaffe
    Beckman DXC 800 CRE 3 kinetic Jaffe
    Roche Modular P-800 creatinine plus enzymatic assay
    Roche Modular P-800 kinetic Jaffe
    Vitros 5.1 enzymatic creatinine assay