Stability of pO2, pCO2 and pH in heparinized whole blood samples: influence of storage temperature with regard to leukocyte count and syringe material.
Schmidt C, Müller-Plathe O
Purpose
The purpose of this paper was to determine the effects of leukocyte concentration, storage container and temperature on blood gas analysis.
Conclusion
The partial pressure oxygen (PO2) and carbon dioxide (PCO2) were generally stable during storage in glass syringes, but the PO2 declined over the course of 60 min when tonometered with 140 mm mercury (Hg) oxygen and stored at 22 degrees C. The deviations in PO2 were larger in specimens with high leukoyte counts, but no other effects of leukocyte counts were observed. The pH decreased with storage at 22 or 4 degrees C, regardless of container, but the changes were smaller in specimens stored at 4 degrees C.
Studies (2)
Purpose
The purpose of this study was to determine the effects of storage container and temperature on blood gas analysis in heparinated blood.
Summary of Findings
The PO2 was not significantly affected by storage of blood tonometered with 86 mm Hg oxygen in glass syringes for 1 h at 4 or 22 degrees C, but PO2 increased significantly when blood tonometered with 140 mm Hg oxygen was stored at 22, but not 4 degrees C in glass syringes. The PO2 increased progressively in blood tonometered with 86 mm Hg oxygen stored at either temperature in plastic syringes and in blood tonometered with 140 mm Hg oxygen stored at 4 degrees C, but was stable in blood tonometered with 140 mm Hg oxygen stored in plastic syringes at 22 degrees C. The PCO2 was unaffected by storage of blood in glass syringes, but PCO2 increased when blood was stored at either temperature in radiometer syringes and decreased insignificantly when blood was stored in Braun syringes. The pH decreased with storage, regardless of temperature or container, but the changes were smaller in specimens stored at 4 degrees C.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
GasClinical chemistry/auto analyzerSmall moleculepHPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components Tonometered with 140 mm Hg oxygenTonometered with 86 mm Hg oxygenStorage Storage duration 0 min30 min60 minStorage Storage temperature 22 degrees C4 degrees CStorage Type of storage container Glass syringePlastic Braun syringePlastic Radiometer syringe
Purpose
The purpose of this study was to determine the effects of leukocyte concentration and storage temperature on blood gas analysis in heparinated blood. Leukocytes from buffy coats were added to tonometered blood before storage in glass syringes.
Summary of Findings
There were no decreases in PO2 when blood was stored at 4 degrees C, regardless of leukocyte concentration, but when blood was stored at 22 degrees C, there was a progressive decrease in PO2 with larger deviations observed in specimens with high leukocyte counts or those tonometered with 140 mm Hg oxygen. Further, when specimens were stored at 4 degrees C, PCO2 did not change significantly, but when specimens were stored at 22 degrees C, PCO2 increased.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
GasClinical chemistry/auto analyzerSmall moleculepHPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components 20 x 10^9/L leukocytes40 x 10^9/L leukocytes60 x 10^9/L leukocytes,Tonometered with 140 mm Hg oxygenTonometered with 86 mm Hg oxygenStorage Storage duration 0 min30 min60 minStorage Storage temperature 22 degrees C4 degrees C