Measurement of prothrombin time in patients on oral anticoagulant therapy: effect of two different evacuated tubes.
D'Angelo G, Villa C
Purpose
The purpose of this paper was to compare prothrombin time (PT) in specimens collected in glass Vacutainer tubes with those collected in Venoject II tubes.
Conclusion
Regardless of whether tubes were used at the end of their shelf life or not, or whether the tubes were kept out of their pouch for 30 days (maximum recommended by manufacturer) or not, PT tests results were always 10-30% longer when Venoject II tubes were used for collection compared to glass Vacutainer tubes.
Studies (1)
Purpose
The purpose of this study was to compare PT in specimens from patients on oral anticoagulant therapy collected in glass Vacutainer tubes with those collected in Venoject II tubes. The effects of using tubes at the end of their shelf life and leaving tubes out of their pouch for an extended time were also investigated.
Summary of Findings
Regardless of whether tubes were used at the end of their shelf life or not, or whether the tubes were kept out of their pouch for 30 days (maximum recommended by manufacturer) or not, PT tests results were always 10-30% longer when Venoject II tubes were used for collection compared to glass Vacutainer tubes.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Glass Vacutainer tubesVenoject II at the end of shelf lifeVenoject II kept out of pouch 30 daysVenoject II kept out of pouch for 30 days and at the end of shelf lifeVenoject II tubes