Measurement of prothrombin time in patients on oral anticoagulant therapy: effect of two different evacuated tubes.

D'Angelo G, Villa C

Publication: Haematologica (1999) • Vol 84 • Page 656-7PMID: 10406912View on PubMed

Purpose

The purpose of this paper was to compare prothrombin time (PT) in specimens collected in glass Vacutainer tubes with those collected in Venoject II tubes.

Conclusion

Regardless of whether tubes were used at the end of their shelf life or not, or whether the tubes were kept out of their pouch for 30 days (maximum recommended by manufacturer) or not, PT tests results were always 10-30% longer when Venoject II tubes were used for collection compared to glass Vacutainer tubes.

Studies (1)

Purpose

The purpose of this study was to compare PT in specimens from patients on oral anticoagulant therapy collected in glass Vacutainer tubes with those collected in Venoject II tubes. The effects of using tubes at the end of their shelf life and leaving tubes out of their pouch for an extended time were also investigated.

Summary of Findings

Regardless of whether tubes were used at the end of their shelf life or not, or whether the tubes were kept out of their pouch for 30 days (maximum recommended by manufacturer) or not, PT tests results were always 10-30% longer when Venoject II tubes were used for collection compared to glass Vacutainer tubes.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    MorphologyHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Glass Vacutainer tubes
    Venoject II at the end of shelf life
    Venoject II kept out of pouch 30 days
    Venoject II kept out of pouch for 30 days and at the end of shelf life
    Venoject II tubes