Differential effect of blood collection tubes on total free fatty acids (FFA) and total triiodothyronine (TT3) concentration: a model for studying interference from tube constituents.

Bowen RA, Vu C, Remaley AT, Hortin GL, Csako G

Publication: Clin Chim Acta (2007) • Vol 378 • Page 181-93PMID: 17234171View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection tube type and collection tube constituents on total free fatty acids (FFA), total triiodothyronine (TT3), and beta-hydroxybutyrate (B-HB) in serum and plasma.

Conclusion

BB-HB measurement was not affected by collection tube type, but FFA and TT3 were significantly higher in serum separator tubes (SST) compared to Vacuette or glass tubes.

Studies (2)

Purpose

The purpose of this study was to determine the effects of collection tube type on FFA, TT3, and B-HB in serum and plasma. Serum and plasma specimens were transferred to plain plastic test tubes and stored at -25 degrees C for up to 1 day before analysis.

Summary of Findings

Compared to average FFA concentrations measured using Vacuette tubes, FFA concentrations were higher using SST or reformulated SST and similar or lower using glass tubes for blood collection. Importantly, serum collected in glass tubes and transferred to the different collection tube types did not show any differences in FFA measurements between tube types. B-HB levels were not significantly different between blood collected in Vacuette tubes, glass tubes, SST, or reformulated SST. Average TT3 concentrations were significantly higher in blood collected in SST and similar or significantly higher in reformulated SST compared to glass or Vacuette tubes. FFA and B-HB concentrations were significantly higher in heparinated plasma than in serum from plain glass tubes, but FAA concentrations were higher in serum from SST and Vacuette tubes than in K2EDTA plasma. No correlations were found between various blood count parameters and tube-dependent FFA concentrations. The authors state that serum angiotensin-1-converting enzyme activity, aldolase activity, protein, lipoprotein, homocysteine, vitamin A, vitamin E, carotene, creatine kinase, lactate dehydrogenase, and cystatin C were not affected by collection tube type.

  • Biospecimens

    FluidBlood
    FluidPlasma
    FluidSerum
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    LipidClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    LipoproteinClinical chemistry/auto analyzer
    Small moleculeHPLC
    Small moleculeColorimetric assay
    Cell count/volumeHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    None
    Potassium EDTA
    Sodium heparin
    Biospecimen AcquisitionType of collection container/solution
    EDTA
    Glass
    Reformulated SST
    SST
    Sodium heparin
    Vacuette
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Whole blood