Storage of plasma in primary plasma separator tubes.
Madira WM, Wilcox AH, Barron JL
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of centrifugation duration (5, 10 or 20 min) and storage with gel in PST on clinical chemistry analytes in plasma. Blood was collected from a single subject, and plasma was stored in the refrigerator after centrifugation. Plasma aliquots were removed and placed in tubes without gel at set intervals until all were analyzed on day 7. Blood was collected from an additional 5 subjects, and after initial centrifugation, plasma was placed into two tubes, one of which had an additional 6.5 mm gel added prior to storage.
Summary of Findings
Storage of plasma in PST with no additional gel for 7 days resulted in highly variable 8-45% decreases in glucose, -4-36% changes in potassium and 4-20% increases phosphate concentrations compared to plasma from the same 6 patients that was immediately removed from PST and stored in separate tubes. The authors report that centrifugation duration had no effect on decreases in glucose or increases in potassium and phosphate in plasma from a single patient. When plasma was stored with an additional 6.5 mm gel, the magnitude of the changes in glucose, potassium and phosphate was reduced, but progressive effects of storage were observed. Sodium, urea, creatinine, chloride, urate, albumin, total protein, total bilirubin, alkaline phosphate, alanine transaminase, aspartate transaminase, gamma glutamyl transferase, cholesterol, triglyceride, creatine kinase, hydroxybutyrate dehydrogenase, iron and iron binding capacity were unaffected by storage, regardless of container.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
Other PreservativeDiagnoses
NormalPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerCarbohydrateClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerLipidClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Centrifugation Multiple durations comparedStorage Storage duration 0 days1 day2 days3 days5 days7 daysStorage Type of storage container In a PST with 6.5 mm gelIn a new tubeIn the original PST (with gel)