Effects of sample storage time, temperature and syringe type on blood gas tensions in samples with high oxygen partial pressures.
Pretto JJ, Rochford PD
Purpose
The purpose of this paper was to determine the effects of storage container and temperature on blood gas analysis.
Conclusion
The partial pressure carbon dioxide (PCO2) increased and the partial pressure oxygen (PO2) and pH decreased with storage, but the magnitude of the changes depended on syringe material and was larger when blood was stored at room temperature rather than on ice.
Studies (1)
Purpose
The purpose of this study was to determine the effects of storage container and temperature on blood gas analysis. Heparinized blood was tonometered with 95% oxygen and 5% carbon dioxide to have a high PO2 before transfer into the storage containers, which were periodically agitated.
Summary of Findings
The PCO2 increased, and the PO2 and pH decreased with storage, and the rate of change was greater when blood was stored at room temperature than in ice water. The PO2 decreased significantly in specimens stored at room temperature for 5 min, regardless of syringe type, but was generally the most stable in specimens stored in glass syringes and least stable in Preza-Pak syringes. The increases in PCO2 were first significant after at least 40 min of storage and were greatest when specimens were stored in glass or Preza-Pak II syringes. The PCO2 decreased slightly in the first 20 min in specimens stored in OS90 syringes, but then remained relatively stable.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
GasRadiometerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 0 min10 min120 min20 min40 min5 min60 min90 minStorage Storage temperature Ice waterRoom temperature (21-23 degrees C)Storage Type of storage container Glass syringePlastic Marquest Quick syringePlastic Radiometer QS90 syringePlastic Terumo Preza-Pak II syringe