Plasma tissue factor antigen levels in capillary whole blood and venous blood: effect of tissue factor on prothrombin time.

Quien ET, Morales E, Cisar LA, Kim HC, Cimini C, Saidi P, Philipp CS

Publication: Am J Hematol (1997) • Vol 55 • Page 193-8PMID: 9257879View on PubMed

Purpose

The purpose of this paper was to determine the effects of measurement method, using capillary versus venous blood, gender, and age on prothrombin time and tissue factor in specimens from healthy controls and patients on anticoagulant therapy.

Conclusion

There was a linear correlation between tissue factor levels and patient age but no relationship with gender. Capillary tissue factor levels were higher than venous tissue factor levels. The portable analyzer measured slightly lower whole blood prothrombin times in capillary specimens compared to venous specimens, but these whole blood prothrombin times were highly correlated with prothrombin times measured in platelet poor plasma.

Studies (1)

Purpose

The purpose of this study was to determine the effects of measuring prothrombin time and tissue factor in capillary versus venous blood specimens from healthy controls and patients on anticoagulant therapy. In addition, the effects of patient age and gender on tissue factor levels were evaluated, and two types of analyzers for the measurement of prothrombin time were compared. Plasma intended for tissue factor measurement was stored at -70 degrees C until analysis. Prothrombin time was measured in fresh platelet poor plasma and in whole blood.

Summary of Findings

After correction for age, there were no differences in venous or capillary tissue factor between patients on anticoagulation therapy and healthy controls. There was a linear correlation between tissue factor levels and patient age but no relationship with gender. Capillary tissue factor levels were an average of 31 pg/mL higher than venous tissue factor levels (p=0.0001). The portable analyzer measured slightly lower whole blood prothrombin times in capillary specimens compared to venous specimens (-0.6 sec, p=0.004), but these whole blood prothrombin times were highly correlated with prothrombin times measured in platelet poor plasma by the ACL 300 Plus analyzer (r=0.98). The differences in capillary and venous whole blood prothrombin time did not correlate with differences in capillary and venous tissue factor levels.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    FrozenNone (Fresh)
  • Diagnoses

    Normal
    Other diagnoses
  • Platforms

    ProteinELISA
    MorphologyHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Arterial thrombosis
    Atrial fibrillation
    Deep venous thrombosis
    Prosthetic heart valve
    PreaquisitionOther drugs
    Anticoagulation with warfarin
    None
    PreaquisitionPatient age
    19-88 years
    PreaquisitionPatient gender
    Female
    Male
    Biospecimen AcquisitionAnatomical location of blood draw
    Capillary
    Vein
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Platelet-poor plasma
    Whole blood
    Hematology/ auto analyzer SpecificTechnology platform
    ACL 300 Plus laser-nephelometric centrifugal analyzer
    Hemochron Jr. Microcoagulation System (portable)