The use of capillary blood for measurements of circulating ferritin.
Pootrakul P, Skikne BS, Cook JD
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of using capillary blood specimens instead of venous blood specimens for the measurement of ferritin in plasma and serum. Plasma and serum specimens were stored at -20 degrees C unless otherwise indicated.
Summary of Findings
Overall, capillary specimens showed higher between- and within-specimen variability than venous specimens with regards to ferritin concentrations. Furthermore, both capillary plasma and serum had higher average ferritin concentrations (3-5%) than venous specimens. The variability of ferritin concentrations was higher in plasma than in serum after frozen storage of small volumes (0.1-0.2 mL), but with no storage, the variability was similar between plasma and serum. The average concentration of ferritin was significantly lower when 1 mL aliquots of serum were stored in 2 mL vials than when 0.1-0.2 mL aliquots of serum were stored in smaller vials.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
FrozenNone (Fresh)Diagnoses
NormalPlatforms
ProteinImmunoradiometric assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anatomical location of blood draw ArmCapillaryHandLeft and right sides comparedVeinBiospecimen Aliquots and Components Aliquot size/volume 0.1-0.2 mL1 mLBiospecimen Aliquots and Components Blood and blood products PlasmaSerumBiospecimen Preservation Type of fixation/preservation FrozenNone (fresh)