Skip to main content

The use of capillary blood for measurements of circulating ferritin.

Pootrakul P, Skikne BS, Cook JD

Publication: Am J Clin Nutr (1983) • Vol 37 • Page 307-10PMID: 6823894View on PubMed

Purpose

The purpose of this paper was to determine the effects of using capillary blood instead of venous blood specimens for the measurement of ferritin in plasma and serum.

Conclusion

Overall, capillary specimens showed higher between- and within-specimen variability and higher ferritin concentrations than venous specimens. The variability of ferritin concentrations was higher in plasma than in serum after frozen storage of small volumes, but with no storage, the variability was similar between plasma and serum. The average concentration of ferritin was significantly lower when 1 mL aliquots of serum were stored in 2 mL vials than when 0.1-0.2 mL aliquots of serum were stored in smaller vials.

Studies (1)

Purpose

The purpose of this study was to determine the effects of using capillary blood specimens instead of venous blood specimens for the measurement of ferritin in plasma and serum. Plasma and serum specimens were stored at -20 degrees C unless otherwise indicated.

Summary of Findings

Overall, capillary specimens showed higher between- and within-specimen variability than venous specimens with regards to ferritin concentrations. Furthermore, both capillary plasma and serum had higher average ferritin concentrations (3-5%) than venous specimens. The variability of ferritin concentrations was higher in plasma than in serum after frozen storage of small volumes (0.1-0.2 mL), but with no storage, the variability was similar between plasma and serum. The average concentration of ferritin was significantly lower when 1 mL aliquots of serum were stored in 2 mL vials than when 0.1-0.2 mL aliquots of serum were stored in smaller vials.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    FrozenNone (Fresh)
  • Diagnoses

    Normal
  • Platforms

    ProteinImmunoradiometric assay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnatomical location of blood draw
    Arm
    Capillary
    Hand
    Left and right sides compared
    Vein
    Biospecimen Aliquots and ComponentsAliquot size/volume
    0.1-0.2 mL
    1 mL
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen PreservationType of fixation/preservation
    Frozen
    None (fresh)