Ferritin concentrations in plasma from capillary (finger prick) blood and venous blood compared.
Mejía LA, Viteri FE
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of type of storage container during centrifugation and using capillary plasma (finger prick) instead of venous plasma to measure ferritin concentrations in adults and children.
Summary of Findings
Although highly correlated, ferritin concentrations were significantly higher in finger prick plasma obtained through centrifugation in a capillary tube than in venous plasma obtained through centrifugation in a standard blood-collection tube (+8.3%, p<0.0180 and +18.6%, p<0.01 for adults and children, respectively). Further, ferritin concentrations were significantly higher in finger prick plasma than in venous plasma when both were obtained by centrifugation in capillary tubes (+3.20 ug/L, p<0.05). Lastly, ferritin concentrations were significantly higher in venous plasma obtained by centrifugation in a capillary tube than that obtained by centrifugation in the standard blood-collection tube (+4.86 ug/L, p<0.01).
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinImmunoradiometric assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Patient age 14-66 months21-49 yearsBiospecimen Acquisition Anatomical location of blood draw CapillaryVeinBiospecimen Acquisition Method of fluid acquisition Finger/heel prick samplingVenipunctureStorage Type of storage container Capillary tubeStandard blood collection tube