Changes in oxygen measurements when whole blood is stored in iced plastic or glass syringes.

Mahoney JJ, Harvey JA, Wong RJ, Van Kessel AL

Publication: Clin Chem (1991) • Vol 37 • Page 1244-8PMID: 1823532View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage of whole blood or plasma in glass or plastic syringes on partial pressure oxygen (pO2) and partial pressure carbon dioxide (pCO2).

Conclusion

There were no significant changes in pO2 when plasma or whole blood was stored in glass syringes in an ice bath for up to 60 minutes, regardless of initial oxygen levels. The pO2 increased during storage of whole blood or plasma in plastic syringes in an ice bath or whole blood in room temperature plastic syringes, but the magnitude of the increase depended on the initial oxygen concentration. The authors report that there were no significant effects of storage of plasma or whole blood in plastic or glass syringes on the pCO2.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage of whole blood or plasma in glass or plastic syringes in an ice bath and storage of whole blood in room temperature plastic syringes on pO2 and pCO2. Plasma and whole blood specimens from 10 volunteers were tonometered with a mixture of oxygen, carbon dioxide and nitrogen prior to storage.

Summary of Findings

There were no significant changes in pO2 when plasma or whole blood was stored in glass syringes in an ice bath for up to 60 minutes, regardless of initial oxygen concentration. The pO2 in whole blood stored in plastic syringes in an ice bath increased, but the magnitude of the increase was greatest when the initial pO2 was highest and the increase was insignificant in specimens that were tonometered to 60 mL/L. Plasma pO2 increased in all specimens stored for 30 min in plastic syringes in an ice bath, but the increase was largest when the initial pO2 was lowest. When whole blood was stored for 30 min in plastic syringes at room temperature, pO2 increased, but the increase was not significant in specimens tonometered to 140 mL/L. The authors report that there were no significant effects of storage of plasma or whole blood in plastic or glass syringes on the pCO2.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    Other PreservativeNone (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    GasClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen components
    100 mL/L oxygen
    140 mL/L oxygen
    60 mL/L oxygen
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Whole blood
    Biospecimen PreservationType of fixation/preservation
    None (fresh)
    Refrigeration
    StorageStorage duration
    0 min
    30 min
    60 min
    90 min
    StorageStorage temperature
    In an ice bath
    Room temperature
    StorageType of storage container
    Glass syringe
    Plastic syringe