Sodium measurement: effects of differing sampling and analytical methods.

Loughrey CM, Hanna EV, McDonnell M, Archbold GP

Publication: Ann Clin Biochem (2006) • Vol 43 • Page 488-93PMID: 17132280View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection and detection method on the measurement of sodium in whole blood and plasma.

Conclusion

Sodium concentrations measured in capillary blood by the Omni autoanalyzer using the direct ion selective electrode (ISE) method and those measured in venous plasma by the Modular autoanalyzer using the indirect ISE method were poorly correlated, with no significant bias. Sodium concentrations measured in capillary blood using the Omni autoanalyzer (direct ISE) were generally lower than those measured in venous blood by the same autoanalyzer. Sodium concentrations measured in venous blood by the Omni autoanalyzer (direct ISE) were generally higher than those measured in venous plasma by the Modular autoanalyzer (indirect ISE).

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection method (capillary versus venous) and detection method (direct versus indirect ISE) on the measurement of sodium in whole blood and plasma. Capillary and venous specimens were obtained within 5 minutes of one another.

Summary of Findings

Sodium concentrations measured in capillary blood by the Omni autoanalyzer using the direct ISE method and those measured in venous plasma by the Modular autoanalyzer using the indirect ISE method were poorly correlated (r=0.365), with no significant bias. Sodium concentrations measured in capillary blood using the Omni autoanalyzer (direct ISE) were generally lower than those measured in venous blood by the same autoanalyzer (r=0.53). Sodium concentrations measured in venous blood by the Omni autoanalyzer (direct ISE) were generally higher than those measured in venous plasma by the Modular autoanalyzer (indirect ISE) (r=0.70). The magnitude of these differences was not associated with total protein or plasma triglyceride levels.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
    Diabetes Type 1
    Diabetes Type 2
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnatomical location of blood draw
    Capillary
    Vein
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Whole blood
    Clinical chemistry/auto analyzer SpecificDetection method
    Direct ion selective electrode
    Indirect ion selective electrode
    Clinical chemistry/auto analyzer SpecificTechnology platform
    POCT Omni S
    Roche Modular ISE 1800 analyser