A comparison of methods of blood withdrawal and sample preparation for potassium measurements.
Hill AB, Nahrwold ML, Noonan D, Northrop P
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of heparin concentration, sample dilution, draw location, tourniquet application, and delayed analysis on potassium measurement in plasma and serum.
Summary of Findings
Draw location, 40 min delays at room temperature prior to analysis (only plasma evaluated), and use of a tourniquet for 3 min (rather than no tourniquet) had no significant effects on the measurement of potassium levels in plasma or serum. However, comparing specimens taken from the same site on the basis of presence (plasma) or absence of heparin (serum) revealed that serum specimens with no heparin had higher concentrations of potassium than plasma specimens with heparin. Further, increasing heparin concentrations and increasing volumes of diluent decreased potassium levels.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Other diagnosesPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Patients requiring cardiopulmonary bypass during surgeryBiospecimen Acquisition Anatomical location of blood draw ArteryCentral veinPeripheral veinBiospecimen Acquisition Anticoagulant HeparinMultiple concentrations evaluatedNoneBiospecimen Acquisition Cannulation 16-gauge cannula in peripheral vein20-gauge cannula in radial arterySwan-Ganz catheter in pulmonary artery via right internal jugular veinBiospecimen Acquisition Method of fluid acquisition Tourniquet times comparedBiospecimen Aliquots and Components Blood and blood products PlasmaSerumStorage Time at room temperature 40 min5 min