Storage of serum or whole blood samples? Effects of time and temperature on 22 serum analytes.
Heins M, Heil W, Withold W
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of storage of serum or whole blood in the dark on the measurement of 22 clinical chemistry analytes in serum from 10 healthy individuals. Blood specimens were centrifuged on the day of analysis.
Summary of Findings
When serum was stored at 9 degrees C, there was an increase in inorganic phosphate and a decrease in LDH after 7 days, but no other changes were noted. Storage of serum at room temperature led to changes in LDL-cholesterol, bilirubin, creatine kinase (CK), aspartate aminotransferase (AST), inorganic phosphate, triacylglycerols (TAG), uric acid, and HDL-cholesterol, with the first changes occurring after 2 days. Storage of whole blood at 9 degrees C resulted in changes in sodium, creatinine, alkaline phosphatase, potassium, LDH, chloride, alpha-amylase, uric acid, inorganic phosphate, free hemoglobin, and magnesium. Storage of whole blood at room temperature resulted in alterations in chloride, LDL-cholesterol, calcium, sodium, alanine aminotransferase (ALT), alkaline phosphatase, gamma-glutamyltransferase (GGT), alpha-amylase, potassium, inorganic phosphate, magnesium, LDH, creatinine, urea,free hemoglobin and HDL-cholesterol, with some changes occurring after only one day. Only cholinesterase was unaffected by storage of whole blood or serum for up to 7 days at room temperature or 9 degrees C.
Biospecimens
FluidBloodFluidSerumPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerLipidClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerProteinSpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Blood and blood products SerumWhole bloodBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Storage duration 0 days1 day2 days3 days4 days7 daysStorage Storage temperature 9 degrees CRoom temperature