Effects of specimen collection, processing, and storage conditions on stability of human immunodeficiency virus type 1 RNA levels in plasma.
Ginocchio CC, Wang XP, Kaplan MH, Mulligan G, Witt D, Romano JW, Cronin M, Carroll R
Purpose
Conclusion
Studies (2)
Purpose
The purpose of this study was to determine the effects of anticoagulant, plasma separator, freeze-thaw cycling and frozen storage on the HIV load in plasma and serum. Lysed specimens were stored at -70 degrees C prior to analysis.
Summary of Findings
The HIV load was lower in serum than in plasma regardless of anticoagulant, but the difference was greater than the 0.3 log variability of the assay in only 6 of the 10 specimens. When specimens were lysed prior to freezing rather than frozen and thawed prior to lysis, serum HIV loads increased (0.073 log average). HIV load decreased when plasma was frozen prior to lysis, but the magnitude of the decrease was smaller for EDTA plasma (average -0.050 log) than ACD (average -0.219) or heparinated-plasma (average -0.240), however the mean difference for each anticoagulant was not significant. When plasma or serum was stored for 6 months at -70 degrees C, HIV load declined, and the mean differences was more than the 0.3 log assay variability in serum or heparinated-plasma, but not in ACD-plasma or EDTA-plasma.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
None (Fresh)FrozenDiagnoses
AIDS/HIV-relatedPlatforms
RNANucleic acid sequence-based amplification (NASBA)Pre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Acid-citrate-dextroseNonePotassium EDTASodium heparinBiospecimen Aliquots and Components Blood and blood products PlasmaSerumBiospecimen Preservation Type of fixation/preservation FrozenNone (fresh)Storage Storage duration 1 month2 months3 months4 months6 months
Purpose
The purpose of this study was to determine the effects of storing whole blood in EDTA vacutainers or cell free plasma in EDTA-PPT on HIV load.
Summary of Findings
Plasma HIV load was unaffected by storage of whole blood in EDTA vacutainers or cell free plasma in EDTA-PPT at room temperature for up to 30 h. Further, plasma HIV load was similar in cell free plasma stored in EDTA-PPT at room temperature and 4 degrees C. Subjecting EDTA-plasma specimens to a single freeze-thaw cycle did not change the HIV load compared to fresh specimens.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
None (Fresh)FrozenOther PreservativeDiagnoses
AIDS/HIV-relatedPlatforms
RNANucleic acid sequence-based amplification (NASBA)Pre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution EDTA Plasma preparation tubeEDTA vacutainerBiospecimen Aliquots and Components Blood and blood products PlasmaWhole bloodBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedBiospecimen Preservation Type of fixation/preservation FrozenNone (fresh)RefrigerationStorage Storage duration 2 h30 h8 hStorage Storage temperature 23 degrees C4 degrees C