Blood gas analysis: effect of air bubbles in syringe and delay in estimation.

Biswas CK, Ramos JM, Agroyannis B, Kerr DN

Publication: Br Med J (Clin Res Ed) (1982) • Vol 285 • Page 1659-60PMID: 6802352View on PubMed

Purpose

The purpose of this paper was to determine the effects of delayed analysis and air bubbles in the syringe on blood gas determination.

Conclusion

The partial pressure oxygen (PO2) was reduced in venous blood stored for more than 30 min at 0 degrees C, or 10 min at 4 or 22 degrees C compared to levels in blood analyzed after 5 min. The partial pressure carbon dioxide (PCO2) and bicarbonate increased and pH decreased when venous blood was stored for 120 min at 0 degrees C compared to venous blood analyzed after 5 min, but no significant effects were observed after shorter delays in processing. When arterial blood was stored at 22 or 0 degrees C, the PO2 decreased within 10 and 60 min, respectively. Blood containing air bubbles or froth had significantly higher PO2 after 2 min and lower PCO2 after 3 min at room temperature compared to levels measured in immediately analyzed specimens.

Studies (2)

Purpose

The purpose of this study was to determine the effects of blood storage temperature and duration on the analysis of venous and arterial blood gas.

Summary of Findings

The P02 was unaffected by storage of venous blood at 0 degrees C for up to 30 min prior to analysis, but it was reduced in venous blood stored for 60 or 120 min at 0 degrees C compared to venous blood analyzed within 5 min. The pCO2, and bicarbonate increased and pH decreased when venous blood was stored for 120 min at 0 degrees C compared to venous blood analyzed after 5 min, but no significant effects were observed after shorter delays in processing. The PO2 was reduced when venous blood was stored for 20 or 30 min at 4 or 22 degrees C compared to blood analyzed after 5 min, but the PCO2 was unaffected by storage at 4 or 22 degrees C for up to 30 min. When arterial blood was stored at 22 or 0 degrees C, the PO2 decreased within 10 and 60 min, respectively, compared to levels in arterial blood analyzed after 5 min. The authors note the decrease in P02 occurred faster in arterial blood than venous blood most likely due to the higher initial levels.

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
    Other diagnoses
  • Platforms

    GasClinical chemistry/auto analyzer
    Small moleculepH
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Renal failure
    Biospecimen AcquisitionAnatomical location of blood draw
    Artery
    Vein
    StorageStorage duration
    10 min
    120 min
    20 min
    30 min
    5 min
    60 min
    StorageStorage temperature
    0 degrees C (ice)
    22 degrees C
    4 degrees C