Rapid thawing of fresh-frozen plasma with radio wave-based thawing technology and effects on coagulation factors during prolonged storage at 4 degrees C.

Boström F, Ekemar L, Olsson D, Egberg N, Lundahl J

Publication: Vox Sang (2009) • Vol 97 • Page 34-8PMID: 19490580View on PubMedSuggested by: ISBER

Purpose

The purpose of this paper was to determine the effects of thawing method and storage at 4 degrees C on factor V (FV), factor VIII (FVIII) and protein S (PS) levels in plasma.

Conclusion

When plasma was thawed using a radio wave thawing device (RTD), the temperature after thaw was higher than when plasma was thawed by heated air technology (HAT), but there were no differences in the levels of FV, FVIII and PS when the two thawing methods were compared. After thawing, FV and FVIII levels were more than 90% of the prefreezing values, but after 2 weeks of storage at 4 degrees C and a second freeze-thaw cycle, levels decreased to 70% and 40% of the prefreezing values, respectively. PS levels declined by approximately 10% after a 2 weeks thaw duration and a second freeze-thaw cycle.

Studies (1)

Purpose

The purpose of this study was to determine the effects of thawing method and storage at 4 degrees C on FV, FVIII and PS levels in plasma. Plasma was aliquoted and frozen at -75 degrees C within 8 h of collection. HAT thawing took 23 min while RTD took 7.5 min. Specimens were collected before freezing, after thawing, and 1 and 2 weeks after thawing at 4 degrees C and then frozen and analyzed within 30 min of second thaw.

Summary of Findings

When plasma was thawed using RTD, the temperature after thaw was higher than when plasma was thawed by HAT, but there were no differences in the levels of FV, FVIII and PS when the two thaw methods were compared. After thawing, FV and FVIII levels were more than 90% of the prefreezing values, but after 2 weeks of storage at 4 degrees C and a second freeze-thaw cycle, levels decreased to 70% and 40% of the prefreezing values, respectively. PS levels declined by approximately 10% after a 2 week thaw duration and an additional freeze-thaw cycle.
  • Biospecimens

    FluidPlasma
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
  • Platforms

    ProteinHematology/ auto analyzer
    ProteinImmunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageFreeze/thaw cycling
    1 cycle
    2 cycles
    StorageThaw duration
    1 week
    2 weeks
    Less than 20 min