Skip to main content

Effect of sample dilutions on arterial blood gas determinations.

Dennis RC, Ng R, Yeston NS, Statland B

Publication: Crit Care Med (1985) • Vol 13 • Page 1067-8PMID: 3933913View on PubMed

Purpose

The purpose of this paper was to determine the effects of incompletely purging indwelling arterial catheters prior to specimen collection on blood gas determination.

Conclusion

To ensure no dilution effects on the hematocrit, pH, and partial pressure oxygen (pO2), at least 8 mL of blood needed to be flushed from the arterial syringe prior to specimen collection. The partial pressure carbon dioxide (pCO2) was significantly different when 8 mL of blood was flushed rather than 10 mL.

Studies (1)

Purpose

The purpose of this study was to determine the effects of incompletely purging indwelling arterial catheters of blood prior to specimen collection on blood gas determination.

Summary of Findings

When 4 mL of blood or less was flushed from the indwelling catheter, the hematocrit, pH, and paCO2 were lower and the pO2 was higher than when 10 mL of blood was flushed from the indwelling catheter prior to specimen collection. When 6 mL of blood was flushed from the catheter, the hematocrit and pCO2 were lower and the pO2 was higher than when 10 mL was flushed, but the pH was similar. When 8 ml rather than 10 mL of blood was flushed prior to specimen collection, the pCO2 was significantly lower (p<0.05).
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    GasClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    Cell count/volumeHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionCannulation
    10 mL flushed
    2 mL flushed
    4 mL flushed
    6 mL flushed
    8 mL flushed
    Unflushed saline heparin