Serum microarrays for large scale screening of protein levels.

Janzi M, Odling J, Pan-Hammarström Q, Sundberg M, Lundeberg J, Uhlén M, Hammarström L, Nilsson P

Publication: Mol Cell Proteomics (2005) • Vol 4 • Page 1942-7PMID: 16131484View on PubMed

Purpose

The purpose of this paper was to determine the effects of analysis method and freeze-thaw cycling on serum immunoglobulin A (IgA) and C3 concentrations.

Conclusion

One freeze-thaw cycle led to significant deviations in IgA measurements. A high level of correlation was observed between serum IgA levels measured by microarray and nephelometric analysis. Only 65 of 1964 specimens (3.3%) showed greater than 5-fold deviation between IgA measurements taken by the two methods, with similar percentages having higher values on the array or by nephelometry. Similar correlations between serum microarray and nephelometric analysis were observed for the measurement of C3.

Studies (1)

Purpose

The purpose of this study was to determine the effects of analysis method and freeze-thaw cycling on serum IgA and C3 concentrations. Specimens were collected between 1999 and 2002 and stored at -20 degrees C.

Summary of Findings

One freeze-thaw cycle led to significant deviations in IgA measurements which were larger for specimens with lower initial IgA concentrations (56% for <0.07 mg/mL IgA) than for specimens with higher initial IgA concentrations (12% for >0.07-4.9 mg/mL IgA and 31% for >5 mg/mL IgA). However, the authors note that the lower concentrations of IgA were initially measured by radial immunodiffusion which has a lower limit of detection than nephelometry which was used for measurement after the freeze-thaw cycle. Nephelometric determination of IgA at high concentrations (>5 mg/mL) showed increased variation, and these specimens were excluded from analysis. The authors state that a high level of correlation was observed between serum microarray and nephelometric analysis of serum IgA levels. 65 of 1964 specimens (3.3%) showed greater than 5-fold deviation between IgA measurements taken by each method. 31 of these specimens had higher values on the array, but very low measurements (median 0.02 mg/mL) and 34 had higher values by nephelometry. Similar correlations between serum microarray and nephelometric analysis were observed for the measurement of C3.
  • Biospecimens

    FluidSerum
  • Preservation Types

    Frozen
  • Diagnoses

    Other diagnoses
    Not specified
  • Platforms

    ProteinReverse phase protein microarray
    ProteinRadial immunodiffusion assay
    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    Increased IgA levels
    Low IgA levels
    Normal IgA levels
    PreaquisitionDiagnosis/ patient condition
    Increased susceptibility to infection
    StorageFreeze/thaw cycling
    0 cycles
    1 cycle
    Reverse phase protein microarray SpecificTargeted peptide/protein
    C3
    IgA
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Radial immunodiffusion assay
    Serum microarray