Effect of heparin on umbilical arterial blood gases.
Kirshon B, Moise KJ Jr
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of dilution and heparin concentration on blood gas analysis in arterial umbilical cord blood. Blood drawn into syringes containing a coating of 1,000 U/mL heparin was used as reference.
Summary of Findings
When the syringe contained 0.2 ml of 1,000 U/mL heparin, the pCO2, bicarbonate, and base excess were lower than when the syringe contained a coating of 1,000 U/mL heparin. Use of a coating of 10,000 rather than 1,000 U/mL heparin resulted in decreased bicarbonate and base excess, and when 0.2 mL 10,000 U/mL heparin was used pH, pCO2, bicarbonate and base excess were decreased, and pO2 was increased compared to when a coating of 1,000 U/mL heparin was used.
Biospecimens
FluidCord BloodPreservation Types
None (Fresh)Diagnoses
PregnantPlatforms
Small moleculeClinical chemistry/auto analyzerGasClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Multiple concentrations evaluatedSodium heparin