Effect of heparin on umbilical arterial blood gases.

Kirshon B, Moise KJ Jr

Publication: J Reprod Med (1989) • Vol 34 • Page 267-9PMID: 2715986View on PubMed

Purpose

The purpose of this paper was to determine the effects of dilution and heparin concentration on blood gas analysis in arterial umbilical cord blood.

Conclusion

When the syringe contained 0.2 ml of 1,000 units (U)/mL heparin, the partial pressure carbon dioxide (pCO2), bicarbonate, and base excess were lower than when the syringe contained a coating of the same concentration of heparin. Use of a coating of 10,000 rather than 1,000 U/mL heparin resulted in decreased bicarbonate and base excess, and when 0.2 mL 10,000 U/mL heparin was used, pH, pCO2, bicarbonate and base excess were decreased, and partial pressure oxygen (pO2) was increased compared to when a coating of 1,000 U/mL heparin was used.

Studies (1)

Purpose

The purpose of this study was to determine the effects of dilution and heparin concentration on blood gas analysis in arterial umbilical cord blood. Blood drawn into syringes containing a coating of 1,000 U/mL heparin was used as reference.

Summary of Findings

When the syringe contained 0.2 ml of 1,000 U/mL heparin, the pCO2, bicarbonate, and base excess were lower than when the syringe contained a coating of 1,000 U/mL heparin. Use of a coating of 10,000 rather than 1,000 U/mL heparin resulted in decreased bicarbonate and base excess, and when 0.2 mL 10,000 U/mL heparin was used pH, pCO2, bicarbonate and base excess were decreased, and pO2 was increased compared to when a coating of 1,000 U/mL heparin was used.

  • Biospecimens

    FluidCord Blood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Pregnant
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Multiple concentrations evaluated