Use of glass capillaries avoids the time changes in high blood PO(2) observed with plastic syringes.

d'Ortho MP, Delclaux C, Zerah F, Herigault R, Adnot S, Harf A

Publication: Chest (2001) • Vol 120 • Page 1651-4PMID: 11713149View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage container, duration and temperature on the partial pressure oxygen (pO2) of blood specimens.

Conclusion

When specimens with a pO2 greater than 200 mm mercury (Hg) were stored in plastic syringes, the pO2 declined with storage, but no declines were seen for specimens with lower initial pO2. Although pO2 declined with storage in glass syringes at room temperature, no declines in pO2 were observed when the specimens were stored at 4 degrees C in glass syringes or capillaries.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage container, duration and temperature on pO2 of blood specimens with a range of initial values. Heparinized blood was equilibrated with the appropriate amount of oxygen at 37 degrees C before storage at various temperatures.

Summary of Findings

When the initial pO2 was more than 200 mm Hg, the pO2 fell linearly during storage at room temperature or 4 degrees C, but the magnitude of the decline depended on the initial pO2 and storage container. When the initial pO2 was 650 mm Hg, the fall in pO2 with storage was greater when the specimen was stored at 4 degrees C or room temperature in plastic syringes rather than glass syringes or glass capillaries. When the initial pO2 was 400 mm Hg, the fall in pO2 with storage was greater when the specimen was stored at room temperature, but not 4 degrees C in plastic syringes rather than a glass syringes or glass capillaries. There were no effects of storage container when the initial pO2 was 200 mm Hg or less. While significant declines in pO2 were observed with storage at room temperature when the initial pO2 was 650 or 400 mm HG, the pO2 did not change when specimens with an initial pO2 of 650 or 400 mm Hg were stored at 4 degrees C in glass syringes or glass capillaries.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    GasClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen components
    130 mm Hg O2
    200 mm Hg O2
    400 mm Hg O2
    650 mm Hg O2
    80 mm Hg O2
    StorageStorage duration
    0 min
    15 min
    30 min
    5 min
    60 min
    StorageStorage temperature
    4 degrees C
    Room temperature
    StorageType of storage container
    3 mL plastic syringe
    5 mL glass syringe
    Glass capillary microsampler (Roche)