The effects of time delay and temperature on capillary blood gas measurements.
Dent RG, Boniface DR, Fyffe J, Yousef Z
Purpose
The purpose of this paper was to determine the effects of storage temperature and duration on the measurement of blood gas levels.
Conclusion
There were only non-significant decreases in partial pressure oxygen (pO2) from baseline values when blood specimens were stored for 1 or 2 h at room temperature or on ice; however, when stored on ice or at room temperature, partial pressure carbon dioxide (pCO2) increased. The changes in pCO2 were smaller when specimens were stored at room temperature than when stored on ice.
Studies (1)
Purpose
The purpose of this study was to determine the effects of storage temperature and duration on blood gas analysis. Arterialized capillary blood specimens were collected into 6 capillary tubes for each of 31 subjects. Specimens were mixed magnetically prior to analysis.
Summary of Findings
There were only non-significant decreases in pO2 from baseline values when blood specimens were stored for 1 or 2 h at room temperature or on ice; however, when stored on ice or at room temperature, pCO2 increased. The changes in pCO2 were smaller when specimens were stored at room temperature than when stored on ice. The authors do not report if there was an effect of storage for 1 versus 2 h.
Biospecimens
FluidBloodPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
ProteinLC-MS or LC-MS/MSPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 0 min1 h2 hStorage Storage temperature On iceRoom temperature