Skip to main content

The effects of time delay and temperature on capillary blood gas measurements.

Dent RG, Boniface DR, Fyffe J, Yousef Z

Publication: Respir Med (1999) • Vol 93 • Page 794-7PMID: 10603628View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage temperature and duration on the measurement of blood gas levels.

Conclusion

There were only non-significant decreases in partial pressure oxygen (pO2) from baseline values when blood specimens were stored for 1 or 2 h at room temperature or on ice; however, when stored on ice or at room temperature, partial pressure carbon dioxide (pCO2) increased. The changes in pCO2 were smaller when specimens were stored at room temperature than when stored on ice.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage temperature and duration on blood gas analysis. Arterialized capillary blood specimens were collected into 6 capillary tubes for each of 31 subjects. Specimens were mixed magnetically prior to analysis.

Summary of Findings

There were only non-significant decreases in pO2 from baseline values when blood specimens were stored for 1 or 2 h at room temperature or on ice; however, when stored on ice or at room temperature, pCO2 increased. The changes in pCO2 were smaller when specimens were stored at room temperature than when stored on ice. The authors do not report if there was an effect of storage for 1 versus 2 h.
  • Biospecimens

    FluidBlood
  • Preservation Types

    Other Preservative
  • Diagnoses

    Not specified
  • Platforms

    ProteinLC-MS or LC-MS/MS
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageStorage duration
    0 min
    1 h
    2 h
    StorageStorage temperature
    On ice
    Room temperature