Standardization of a sample preparation and analytical workflow for proteomics of archival endometrial cancer tissue.
Alkhas A, Hood BL, Oliver K, Teng PN, Oliver J, Mitchell D, Hamilton CA, Maxwell GL, Conrads TP
Purpose
The purpose of this paper was to determine the effects of trypsin concentration, the volume of laser microdissected (LMD) tissue, and peptide load analyzed on the number of peptide and protein identifications obtained through liquid chromatography tandem mass spectrometry (LC-MS/MS).
Conclusion
Trypsin to total protein ratios of 1:10, 1:20, and 1:50 resulted in equivalent numbers of identified peptides and proteins. The number of identified peptides increased linearly with the amount of LMD tissue captured, and there were no differences between peptide amounts identified in endometrial stroma or malignant epithelium. Likewise, analysis of increasing total peptide load amounts from 100 up to 1000 ng resulted in an increased number of identified peptides.
Studies (1)
Purpose
The purpose of this study was to determine the effects of trypsin concentration used for protein digestion, the volume of LMD tissue, and the peptide load analyzed on the number of peptide and protein identifications obtained through LC-MS/MS. LMD endometrial tissue specimens were stored in distilled water at -80 degrees C until analysis.
Summary of Findings
The authors state that trypsin to total protein ratios above 1:50 resulted in some intense, broadly eluting peaks but that 1:10, 1:20, and 1:50 ratios resulted in equivalent numbers of identified peptides and proteins. The number of identified peptides increased linearly with the amount of LMD tissue captured, and there were no differences between peptide amounts identified in endometrial stroma or malignant epithelium. Likewise, analysis of increasing total peptide load amounts from 100 to 1000 ng resulted in an increased number of identified peptides, but the authors state that when 2000 ng peptides was used as a load volume, the number of identified peptides decreased.
Biospecimens
TissueEndometriumPreservation Types
FormalinDiagnoses
NeoplasticCarcinomaNeoplasticNormal AdjacentPlatforms
PeptideLC-MS or LC-MS/MSPeptideSpectrophotometryProteinLC-MS or LC-MS/MSMorphologyLight microscopyPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Biospecimen location Endometrial stromaMalignant epitheliumBiospecimen Aliquots and Components Aliquot size/volume 100 ng peptides1000 ng peptides2000 ng peptides250 ng peptides500 ng peptidesBiospecimen Aliquots and Components Cell number 31.5 nL tissue63 nL tissue7 nL tissueAnalyte Extraction and Purification Protein digestion 1:10 trypsin to total protein1:20 trypsin to total protein1:50 trypsin to total protein