Skip to main content

Standardization of a sample preparation and analytical workflow for proteomics of archival endometrial cancer tissue.

Alkhas A, Hood BL, Oliver K, Teng PN, Oliver J, Mitchell D, Hamilton CA, Maxwell GL, Conrads TP

Publication: J Proteome Res (2011) • Vol 10 • Page 5264-71PMID: 21932769View on PubMed

Purpose

The purpose of this paper was to determine the effects of trypsin concentration, the volume of laser microdissected (LMD) tissue, and peptide load analyzed on the number of peptide and protein identifications obtained through liquid chromatography tandem mass spectrometry (LC-MS/MS).

Conclusion

Trypsin to total protein ratios of 1:10, 1:20, and 1:50 resulted in equivalent numbers of identified peptides and proteins. The number of identified peptides increased linearly with the amount of LMD tissue captured, and there were no differences between peptide amounts identified in endometrial stroma or malignant epithelium. Likewise, analysis of increasing total peptide load amounts from 100 up to 1000 ng resulted in an increased number of identified peptides.

Studies (1)

Purpose

The purpose of this study was to determine the effects of trypsin concentration used for protein digestion, the volume of LMD tissue, and the peptide load analyzed on the number of peptide and protein identifications obtained through LC-MS/MS. LMD endometrial tissue specimens were stored in distilled water at -80 degrees C until analysis.

Summary of Findings

The authors state that trypsin to total protein ratios above 1:50 resulted in some intense, broadly eluting peaks but that 1:10, 1:20, and 1:50 ratios resulted in equivalent numbers of identified peptides and proteins. The number of identified peptides increased linearly with the amount of LMD tissue captured, and there were no differences between peptide amounts identified in endometrial stroma or malignant epithelium. Likewise, analysis of increasing total peptide load amounts from 100 to 1000 ng resulted in an increased number of identified peptides, but the authors state that when 2000 ng peptides was used as a load volume, the number of identified peptides decreased.
  • Biospecimens

    TissueEndometrium
  • Preservation Types

    Formalin
  • Diagnoses

    NeoplasticCarcinoma
    NeoplasticNormal Adjacent
  • Platforms

    PeptideLC-MS or LC-MS/MS
    PeptideSpectrophotometry
    ProteinLC-MS or LC-MS/MS
    MorphologyLight microscopy
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionBiospecimen location
    Endometrial stroma
    Malignant epithelium
    Biospecimen Aliquots and ComponentsAliquot size/volume
    100 ng peptides
    1000 ng peptides
    2000 ng peptides
    250 ng peptides
    500 ng peptides
    Biospecimen Aliquots and ComponentsCell number
    31.5 nL tissue
    63 nL tissue
    7 nL tissue
    Analyte Extraction and PurificationProtein digestion
    1:10 trypsin to total protein
    1:20 trypsin to total protein