Pre-analytical operating procedures for serum Low Molecular Weight protein profiling.
Di Girolamo F, Alessandroni J, Somma P, Guadagni F
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of clotting time, storage temperature and duration, the addition of protease inhibitor, and manual versus automated peptide purification on low molecular weight serum proteome profiling. The effects of the addition of protease inhibitor to blood on factor II and fibrinogen were also evaluated.
Summary of Findings
The addition of protease inhibitor had no effects on the coagulation cascade and all serum specimens were hemolysis-free. Use of automated magnetic beads purification resulted in the identification of 68 peaks with a coefficient of variation (CV) of 14.1% while use of manual magnetic beads purification lead to the identification of 57 spectral peaks with a CV of 18.2%. More spectral peaks were recovered when blood was left at room temperature to clot for 1-2 h as opposed to 30 min or 4 h prior to serum analysis. When whole blood was treated with protease inhibitor, storage of serum for up to 2 h at room temperature or 3 h at 4 degrees C prior to analysis resulted in minimal changes to serum proteomes compared to unstored serum, but longer storage at these temperatures resulted in considerable changes to the serum proteomes. When no protease inhibitor was added to whole blood, minimal changes to the serum proteomes were observed after storage of serum at room temperature or 4 degrees C for 1 h, but longer storage times at either temperature resulted in substantial changes to the proteome profiles. Storage of serum at -80 degrees C for 1 month resulted in only minimal changes to the proteome profiles compared to those observed for fresh serum.
Biospecimens
FluidBloodFluidSerumPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
PeptideMALDI-TOF MSProteinMALDI-TOF MSProteinHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Storage duration 1 h1 month12 h2 h24 h3 h6 hStorage Storage temperature -80 degrees C4 degrees CRoom temperatureStorage Time at room temperature 1 h2 h30 min4 hAnalyte Extraction and Purification Analyte purification Automated magnetic beads purificationManual magnetic beads purificationAnalyte Extraction and Purification Protease inhibitor EDTA-freeNo protease inhibitor added