Methodological aspects on microdialysis protein sampling and quantification in biological fluids: an in vitro study on human ventricular CSF.
Dahlin AP, Wetterhall M, Caldwell KD, Larsson A, Bergquist J, Hillered L, Hjort K
Purpose
The purpose of this paper was to determine the effects of perfusion fluid components and modification of microdialysis catheters on fluid recovery, protein adsorption and protein identification by isobaric tags for relative and absolute quantification (iTRAQ) labeling followed by MALDI-TOF mass spectrometry using cerebrospinal fluid (CSF).
Conclusion
Use of surface modified catheters with dextran 500 for microdialysis of CSF consistently resulted in close to 100% fluid recovery, less protein adsorption, and higher recoveries of hemopexin, beta hemoglobin, and transferrin.
Studies (1)
Purpose
The purpose of this study was to determine the effects of perfusion fluid components and modification of microdialysis catheters on fluid recovery, protein adsorption and protein identification by iTRAQ labeling followed by MALDI-TOF mass spectrometry in CSF from a patient with severe spontaneous subarachnoid hemorrhage. The effects were studied in vitro using CSF collected by ventricular drainage and stored frozen at -70 degrees C. Prior to microdialysis, the CSF was heated to 37 degrees C and equilibrated for 30 min.
Summary of Findings
When no dextran was added to the perfusion fluid, only 68% of the fluid was recovered through the surface modified catheter, and no fluid was recovered through the unmodified catheter. When the perfusion fluid contained 2-4% dextran 60, the fluid recovery was greater than 100%, regardless of catheter treatment, but addition of 3% dextran 250 or 500 resulted in superior fluid recovery rates of 105% and 99% for the untreated catheter and 89% and 98% for the surface modified catheter. The transmembrane flux and protein adsorption was lower when surface modified catheters were used for microdialysis than when untreated catheters were used, but the difference in transmembrane flux was only significant when the perfusion fluid contained dextran 60 or dextran 250. Importantly, the variability in fluid recovery between specimens was lower when dialyzed through a surface modified catheter. 48 proteins were identified by iTRAQ of which 16 were present across all preanalytical conditions. Albumin recovery was unaffected by dextran size, or catheter type. Clusterin and complement C3 recovery were not affected significantly by catheter modification. In contrast, recovery of hemopexin, beta hemoglobin, and transferrin were higher when microdialysis was with a surface modified catheter rather than an untreated catheter.
Biospecimens
FluidCerebrospinal FluidPreservation Types
FrozenDiagnoses
Other diagnosesPlatforms
PeptideMALDI-TOF MSProteinMALDI-TOF MSCell count/volumeMacroscopic observationPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Severe spontaneous subarachnoid hemorrhageBiospecimen Acquisition Type of collection container/solution Pluronic F-127 surface modified catheterUntreated catheterBiospecimen Aliquots and Components Filtration 2%3%4%Dextran 250Dextran 500Dextran 60MALDI-TOF MS Specific Targeted peptide/protein AlbuminBeta hemoglobinClusterinComplement C3HemopexinTransferrin