Recovery of microarray-quality RNA from frozen EDTA blood samples.
Beekman JM, Reischl J, Henderson D, Bauer D, Ternes R, Peña C, Lathia C, Heubach JF
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of delayed freezing, processing, and extraction methods on RNA quantity and quality and microarray analysis of EDTA whole blood. RNA obtained from EDTA blood that was frozen immediately after collection, thawed 2 h on ice, and extracted using the QIAamp kit (standard protocol) or PAXgene kit (protocol A), or obtained from EDTA blood that was stored at room temperature for 3 h prior to freezing, and processed like protocol A (protocol B) was compared to RNA obtained from blood collected into a Paxgene tube, frozen immediately, and thawed for 16-21 h at room temperature prior to extraction using the Paxgene kit (reference protocol).
Summary of Findings
When RNA was extracted using the standard protocol, the average yield was 0.07 ug/mL, while an average of 2.84 ug/mL was obtained using the reference protocol. This low yield was not always sufficient, even when pooled, to determine the RNA integrity number (RIN) or to use for microarray analysis. Average RNA yields of 1.76 and 1.29 ug/mL were obtained using protocols A and B, respectively. RINs were lower in RNA obtained by protocols A (6.1) and B (6.4) than RNA obtained using the reference protocol (9.8). When RNA was extracted using the reference protocol, the lowest GAPDH 3 prime to 5 prime ratios, scaling factor and interindividual variability, and the highest percent present calls were obtained. In contrast, when RNA was extracted using the standard protocol, the highest GAPDH 3 prime to 5 prime ratios, and scaling factor, and the lowest percent present calls were obtained. RNA obtained by protocols A and B performed similarly, and was more like RNA obtained by the reference protocol than like RNA obtained by the standard protocol. The call concordance percentages between the specimens obtained using the reference protocol and either the standard protocol, protocol A or protocol B were 77.1%, 87.5% and 85.2%, respectively.
Biospecimens
FluidBloodPreservation Types
FrozenPAXgeneDiagnoses
NormalPlatforms
RNASpectrophotometryRNAAutomated electrophoresis/BioanalyzerRNADNA microarrayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Paxgene tubePotassium EDTA tubeBiospecimen Preservation Type of fixation/preservation FrozenPAXgeneStorage Thaw duration 16-21 h2 hStorage Thaw temperature/condition On iceRoom temperatureStorage Time at room temperature 0 h3 hAnalyte Extraction and Purification Analyte isolation method PAXgene kitQIAamp kit