Accuracy and precision of a new, portable, handheld blood gas analyzer, the IRMA.
Wahr JA, Lau W, Tremper KK, Hallock L, Smith K
Purpose
The purpose of this paper was to determine the effects of autoanalyzer type on pH and blood gas measurements.
Conclusion
There was a high degree of correlation between pH, partial pressure oxygen (PO2) and partial pressure carbon dioxide (PCO2) levels measured with each of three analyzers, but there was a larger negative bias in PCO2 and positive bias in PO2 using the IRMA rather than ABL-50 analyzer compared to the NOVA analyzer for specimens with a PCO2 outside the 30-58 mM mercury (Hg) range or a PO2 of less than 100 mM Hg. When the PO2 was greater than 100 mm Hg, there was more variability, and there was a similar bias using the ABL-50 and IRMA analyzers compared to the NOVA analyzer. The authors report no effects of storage temperature or hematocrit on the measurements of PCO2 or PO2, regardless of autoanalyzer.
Studies (1)
Purpose
The purpose of this study was to determine the effects of temperature, hematocrit and autoanalyzer on pH and blood gas measurements in specimens from 25 patients undergoing cardiovascular bypass surgery.
Summary of Findings
There was a high degree of correlation between pH (all r>0.87), PO2 (all r =0.99) and PCO2 (all r>0.91) levels measured with each of the analyzers. When compared to values obtained using the NOVA analyzer, the bias in pH measurements was similar between the ABL-50 and IRMA analyzers, and while the standard deviation was higher using the IRMA, the authors report the precision was sufficient for clinical use. When the PCO2 was between 30 and 58 mM Hg, the bias and precision were comparable between analyzers, but more specimens fell outside this range using the IRMA than the ABL-50 and exhibited a larger negative bias compared to the NOVA analyzer. However, the percentage of specimens falling outside the range of good agreement was low, and the authors report that it is unlikely to be clinically important. When the PO2 was less than 100 mM Hg, there was a positive bias of 5 mM Hg when measured using the IRMA analyzer compared to the NOVA analyzer, but there was no significant bias using the ABL-50 analyzer compared to the NOVA analyzer. When the PO2 was more than 100 mM Hg, the biases using the ABL and IRMA analyzers, compared to the NOVA analyzer, were 7.4 and 7.6 mM Hg, respectively. Regardless of analyzer used, the variability was higher when the PO2 was more than 100 mM Hg. The authors report no effects of storage temperature or hematocrit on the measurements of PCO2 or PO2, regardless of autoanalyzer.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
GasClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components 11-44% red blood cellsStorage Storage temperature 25.5-40 degrees CClinical chemistry/auto analyzer Specific Technology platform ABL-50IRMANOVA Statlabs Profile 5