Use of syringes containing dry (lyophilized) heparin in sampling blood for pH measurement and blood-gas analysis.
Madiedo G, Sciacca R, Hause L, Sasse E
Purpose
The purpose of this paper was to determine the effects of collecting blood in syringes containing lyophilized heparin on blood gas measurements.
Conclusion
There was no effect of collecting blood in a syringe containing lyophilized heparin rather than a small volume of liquid heparin on pH, partial pressure carbon dioxide (PCO2) and partial pressure oxygen (PO2). When blood was collected in syringes containing liquid heparin was transferred to syringes containing lyophilized heparin, neither specimen volume nor syringe type had any effects on pH, PO2 and PCO2.
Studies (1)
Purpose
The purpose of this study was to determine the effects of collecting different volumes of arterial blood in syringes containing lyophilized heparin on pH, PO2 and PCO2 in specimens from critical care patients.
Summary of Findings
When blood was collected in 3 mL disposable Omnistik syringes containing lyophilized heparin, the pH, PO2 and PCO2 were not significantly different than when the blood was collected in 6 mL Monoject syringes containing 0.2 mL liquid heparin. When aliquots of blood collected in Monoject syringes containing liquid heparin were transferred to Omnistik (0.5, 1, 2, and 3 mL) and Ministik (0.2, 0.5 and 1 mL) syringes containing lyophilized heparin, pH, pO2 and pCO2 were unaffected by syringe type or volume.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
GasClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Monoject syringe containing 0.2 mL liquid heparinOmnistik syringe containing lyophilized heparinBiospecimen Aliquots and Components Aliquot size/volume 0.2 mL0.5 mL1 mL2 mL3 mLStorage Type of storage container Ministik syringeOmnistik syringe