Skip to main content

Effect of heparin on arterial blood gases.

Ordog GJ, Wasserberger J, Balasubramaniam S

Publication: Ann Emerg Med (1985) • Vol 14 • Page 233-8PMID: 3919622View on PubMed

Purpose

The purpose of this paper was to determine the effects of heparin on arterial blood gas analysis.

Conclusion

Partial pressure carbon dioxide (PCO2), partial pressure oxygen (PO2), and calculated bicarbonate (HCO3) levels decreased while base excess increased with increasing volumes of heparin in the syringe, but specimen pH was unaffected. When blood gas values in the absence of heparin were extrapolated from the dose-response curve, levels were not significantly different than in specimens collected with a coating (0.025 mL) of liquid heparin.

Studies (1)

Purpose

The purpose of this study was to determine the effects of heparin on arterial blood gas analysis. From 50 patients, 2 mL of blood was drawn from a single radial artery puncture into three syringes containing different volumes of Ivenex liquid heparin. Syringes were transported to the laboratory on ice and analyzed within 15 min.

Summary of Findings

PCO2, PO2, and calculated HCO3 levels decreased (all p<0.05) while the base excess increased (p<0.05) with increasing volumes of heparin in the syringe, but specimen pH was unaffected. The authors state the changes observed were far greater than expected due to dilution effect alone. Importantly, blood from all 50 patients showed similar changes in blood gas measurements with increasing heparin concentration. When blood gas values in the absence of heparin were extrapolated from the dose-response curve, levels were not significantly different than in specimens collected with a coating (0.025 mL) of liquid heparin.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Chronic Obstructive Pulmonary Disease (COPD)
    Cardiovascular Disease
    NeoplasticCarcinoma
    Not specified
    Pneumonia/Respiratory Infection
    Asthma
    Lupus
    Other diagnoses
  • Platforms

    GasClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    Small moleculepH
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Heparin
    Multiple concentrations evaluated
    Biospecimen Aliquots and ComponentsBiospecimen components
    0.5 mL liquid heparin
    1.0 mL liquid heparin
    Coating of liquid heparin (0.025 mL)