Skip to main content

Studies of the stability of 18 chemical constituents of human serum.

Wilson SS, Guillan RA, Hocker EV

Publication: Clin Chem (1972) • Vol 18 • Page 1498-503PMID: 4639860View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage temperature and duration on the levels of clinical chemistry analytes in serum.

Conclusion

Potassium, sodium, chloride, albumin and urea nitrogen were the only analytes in this study unaffected by storage for up to 42 days at -10, 4, 23-27, or 37 degrees C. The direction of change for all other analytes was temperature, storage duration and analyte-dependent.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage temperature and duration on the levels of clinical chemistry analytes in serum. Leftover serum was pooled and frozen at the end of each collection day and used for this study.

Summary of Findings

Potassium, sodium, chloride, albumin and urea nitrogen were the only analytes unaffected by storage for up to 42 days at -10, 4, 23-27, or 37 degrees C. The direction of change for all other analytes was temperature, duration and analyte-dependent. Calcium and phosphorous were altered by storage of serum for 34 days or more at -10 degrees C, but all other analytes were unaffected by storage of serum at this temperature. Alkaline phosphatase and aspartate aminotransferase (AST) were stable at -10 degrees C for the entire 42 days but were unstable at all other temperatures. Bilirubin (direct and total), glucose, cholesterol, and lactate dehydrogenase (LD) were affected by storage of serum for any time at 37 degrees C, but were slightly more stable in serum stored at 4 degrees C or room temperature. Creatinine and total protein levels were only affected by storage of serum at 37 degrees C for more than 14 or 25 days, respectively. Carbon dioxide, and uric acid were affected by storage of serum at 4 degrees C for more than 12 and 6 days, respectively, or storage of serum at 37 degrees C for more than 14 and 18 days, respectively, but only carbon dioxide was affected by storage at room temperature.
  • Biospecimens

    FluidSerum
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageStorage duration
    0-42 days
    StorageStorage temperature
    -10 degrees C
    23-27 degrees C
    37 degrees C
    4 degrees C