Long- and short-term in vitro D-dimer stability measured with INNOVANCE D-Dimer.

Böhm-Weigert M, Wissel T, Muth H, Kemkes-Matthes B, Peetz D

Publication: Thromb Haemost (2010) • Vol 103 • Page 461-5PMID: 20126827View on PubMed

Purpose

The purpose of this paper was to determine the effects of storing and freezing plasma specimens on D-dimer stability.

Conclusion

D-dimer remained stable when plasma was stored in the original collection tube at room temperature for 4 hours after centrifugation and after an additional 24 h at 2-8 degrees C. Up to 4 freeze-thaw cycles over the course of 14 days did not significantly affect D-dimer stability in citrated plasma. A small decline in D-dimer values was observed after freezing but the decline was not time dependent and, in most cases, was clinically irrelevant.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage at room temperature, refrigerated storage, long-term frozen storage, and freeze-thaw cycling on D-dimer stability in plasma specimens with a range of normal and pathological D-dimer levels. Blood specimens were centrifuged within 2 hours of collection, stored in original tubes for an additional 4 hours at room temperature, and then transferred to 2-8 degrees C for 24 h. Alternatively, plasma specimens were transferred to new tubes after centrifugation and frozen at -60 degrees C or below for the evaluation of freeze-thaw cycling effects (4 cycles over 14 days) and long-term stability.

Summary of Findings

D-dimer remained stable when plasma was stored in the original collection tube at room temperature for 4 hours after centrifugation and after an additional 24 h of storage at 2-8 degrees C. Up to 4 freeze thaw cycles over the course of 14 days did not significantly affect D-dimer stability in citrated plasma, compared to measurements taken within 8 hours of collection. While one out of 49 specimens would have shifted classification from possible deep vein thrombosis (DVT) or pulmonary embolism (PE) (0.47-0.49 mg/L fibrinogen equivalent units (FEU)) to normal (0.51 mg/L FEU) after freeze-thaw cycling, this variation was within assay imprecision. D-dimer decreased by an average of 5.1%, 11.7%, 4.8%, and 9.3% after 14 h, 19 months, 25 months, and 36 months of frozen storage, respectively, when baseline measurements were obtained within 12 hours of blood draw. The authors determined a clinically relevant change in D-dimer to be a deviation of greater than 10% from baseline, thus after 19 months of storage, the average change was clinically relevant, however with further storage the clinical relevance was not present. The authors conclude that there was a small decline in D-dimer values after freezing but the decline was not time dependent.

  • Biospecimens

    FluidPlasma
  • Preservation Types

    FrozenNone (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    PeptideHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    < 0.5 mg/L FEU
    > 0.5 mg/L FEU
    Biospecimen PreservationType of fixation/preservation
    Frozen
    None (fresh)
    StorageFreeze/thaw cycling
    0 cycles
    1 cycle
    2 cycles
    3 cycles
    4 cycles
    StorageStorage duration
    14 h
    19 months
    24 h
    25 months
    36 months
    StorageStorage temperature
    -60 degrees C or lower
    2-8 degrees C
    StorageTime at room temperature
    4-6 h
    Less than 12 h
    Less than 2 h
    Less than 8 h
    StorageType of storage container
    New tube
    Original collection tube