A multiplex endpoint RT-PCR assay for quality assessment of RNA extracted from formalin-fixed paraffin-embedded tissues.

Takano EA, Mikeska T, Dobrovic A, Byrne DJ, Fox SB

Publication: BMC Biotechnol (2010) • Vol 10 • Page 89PMID: 21162754View on PubMed

Purpose

The purpose of this paper was to determine if multiplex RT-PCR is suitable to assess RNA integrity of formalin-fixed paraffin-embedded (FFPE) specimens.

Conclusion

RNA yield was higher, but RT-PCR and real-time PCR success rates were lower when RNA was extracted from FFPE specimens with the Trizol method rather than the High Pure kit. Amplification success in the multiplex assay decreased with fragment size. The highest success rates for all three products were in specimens 8 years old rather than 3 or 13 years old. Amplification of the 92 bp product in the multiplex assay was correlated with successful real-time PCR amplification. Maximum multiplex PCR fragment size was inversely related to the real-time PCR cycle threshold (CT).

Studies (1)

Purpose

The purpose of this study was to determine the effects of extraction method, storage duration, and PCR fragment size and platform on the amplification of TATA box binding protein (TBP) RNA from FFPE breast carcinoma specimens. Tumor cells were microdissected from 1-5 freshly cut, deparaffinized, methyl green-stained sections, and 3 replicate extractions were performed with each method.

Summary of Findings

The average RNA yield was slightly higher when RNA was extracted from FFPE tissue using the Trizol method (2.76 ug) compared to the High Pure kit (2.23 ug). The average absorbance 260/280 ratios were 1.7 and 1.8, respectively, for RNA extracted with the Trizol method and the High Pure kit. While the 300 bp fragment of TBP was not amplifiable in any of the specimens, amplification of the 92, 161 and 252 bp products was possible in at least one specimen 3, 8 and 13 years old, but amplification success rates decreased with increasing fragment size. The highest rates of successful amplification for all three products were in specimens 8 years old rather than 3 or 13 years old. More specimens extracted using the High Pure kit had CT values below 35 than specimens extracted using the Trizol method (78 and 72 of 90, respectively). Lack of successful amplification of the 92 bp fragment in the multiplex RT-PCR assay was associated with real-time PCR CT values above 35, but in 135 of 139 specimens from which there was successful amplification of the 92 bp fragment, a CT of less than 35 was obtained by real-time PCR. Further, the average real-time PCR CT decreased as the maximum amplicon size obtained in multiplex PCR increased.
  • Biospecimens

    TissueBreast
  • Preservation Types

    Formalin
  • Diagnoses

    NeoplasticCarcinoma
  • Platforms

    RNAReal-time qRT-PCR
    RNART-PCR
    RNASpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageStorage duration
    13 years
    3 years
    8 years
    Analyte Extraction and PurificationAnalyte isolation method
    High Pure RNA
    Lysis, proteinase k digestion, Trizol extraction and DNAse treatment
    RT-PCR SpecificLength of gene fragment
    161 bp
    252 bp
    92 bp
    RT-PCR SpecificTargeted nucleic acid
    TBP
    RT-PCR SpecificTechnology platform
    Real-time PCR
    Real-time qRT-PCR SpecificTargeted nucleic acid
    TBP