Pre-analytical effects of different lithium heparin plasma separation tubes in the routine clinical chemistry laboratory.

Brandhorst G, Engelmayer J, Götze S, Oellerich M, von Ahsen N

Publication: Clin Chem Lab Med (2011) • Vol 49 • Page 1473-7PMID: 21605014View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection tube type and storage duration on the measurement of 15 different clinical chemistry analytes in plasma.

Conclusion

Almost all analytes measured in this study showed significant changes after storage in plasma separation tubes, regardless of tube manufacturer. Glucose (GLUC), potassium (K), and lactate dehydrogenase (LDH) showed the largest, most clinically significant changes when plasma was stored in the original collection tube after centrifugation. For almost all parameters, significant differences were seen between storage in the 4 plasma separation tubes and in the S-Monovette lithium heparin tubes with a manually installed valve separator and no gel separator, the latter showing smaller changes in analyte levels over time.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection tube type and storage duration on the measurement of 15 different clinical chemistry analytes in plasma. After centrifugation, specimens were stored at 4 degrees in the original collection tubes until analysis at the indicated time points.

Summary of Findings

Almost all analytes measured in this study showed significant changes after storage in plasma separation tubes, regardless of tube manufacturer. In general, aspartate aminotransferase, creatinine, hemolysis index, LDH, lipemia index, K, sodium, and chloride increased with storage duration and alkaline phosphatase, total bilirubin, creatine kinase, and GLUC decreased. Blood urea nitrogen showed little change, and gamma-glutamyltransferase and alanine aminotransferase showed variable changes with storage time that were dependant on collection tube type. GLUC, K, and LDH showed the largest, most clinically significant changes with storage after centrifugation. For almost all parameters, significant differences were seen between storage in the 4 plasma separation tubes and in the S-Monovette lithium heparin tubes with a manually installed valve separator and no gel separator, the latter showing smaller changes in analyte levels over time.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    ProteinClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    Cell count/volumeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    BD Vacutainer lithium heparin PST II gel
    Primavette V sepa lithium heparin rubber separation system
    S-Monovette lithium heparin gel
    S-Monovette lithium heparin with manually installed valve separator
    Vacuette lithium heparin gel
    StorageStorage duration
    24 h
    72 h
    Less than 30 min